S E C T I O N 5
18
RDO Pro-X Optical Dissolved Oxygen Measurement System
|
5
Care and Maintenance
Cleaning the Sensor Cap
1. Leave the cap and nose cone on the sensor!
2. Rinse the sensor with clean water from a squirt bottle or spray bottle.
3. Gently wipe with a soft-bristled brush or soft cloth if biofouling is present. Use Alconox
remove grease.
4. If extensive fouling or mineral build-up is present, soak the cap end in vinegar for 15 min.,
then soak in deionized water for 15 min.
Do not use organic solvents—they will damage the foil.
Do not remove the cap from the sensor prior to brushing.
5. After cleaning the sensor, perform a 1-point user calibration or calibration check.
Cleaning the Optical Window
Perform only when changing the cap. See full instructions in the sensor replacement
cap kit.
Do not wet the lens area with water or any solution.
Remove the cap and gently wipe the window with the supplied lens wipe.
Use only the supplied lens wipe for cleaning. Do not use any other wipe or material.
®
to
Thermo Scientific AquaSensors