Section 2 Preparation for Use
Figure 2-4
Edit Method Settings screen
Figure 2-5

Time-based Fractions screen

2-4
8. Touch the Time icon to display a screen at which you can
modify the fraction collection based on time.
9. The fraction size option should be selected. That is, a box
should appear around the tube. If not, touch the Volume
Select icon to toggle between the options.
10. Use the Arrow icons to set the interval to two seconds.
11. Touch the Enter icon twice to save the setting and return
to the main menu.
__h
__m
_2s
12. Touch the Play icon to start operation. The arm and drop
former will move to the first tube position and then
advance to the next tube every two seconds. Allow the drop
former to advance through at least 20 tubes to ensure that
the arm freely moves through the X and Y axes.
13. Touch the Pause icon. The arm and drop former should
stop over the current tube.
14. Touch the Stop icon. The arm and drop former will move to
the home position.
This completes the preliminary checkout procedure.