Measuring Dissolved Oxygen; General Probe Operation; Measurement - Hach sension6 Instruction Manual

Portable dissolved oxygen meter
Hide thumbs Also See for sension6:
Table of Contents

Advertisement

SECTION 3,
continued

3.4 Measuring Dissolved Oxygen

3.4.1 General Probe Operation

3.4.2 Measurement

Follow the procedures presented below to obtain maximum
performance and accuracy from your sens
Use the DO probe for aqueous applications only.
Take extra care when handling and storing the oxygen
membrane module cap.
Do not allow the DO probe's sensing area (cap reservoir) to
dry out.
Perform the calibration procedure at the beginning of each
day for maximum performance. Recalibrate the DO probe
every two hours for maximum accuracy.
The sample must have a high flow rate or must be stirred
rapidly to obtain accurate results.
Be sure any air bubbles trapped on the probe tip are dislodged
before taking a reading.
It is important to have the DO probe at a uniform
temperature. Do not touch the metallic button on the side of
the probe when holding it. The metallic button is a thermistor
that senses sample temperature. An inaccurate measurement
will result if the temperature of the thermistor is not the same
as the membrane end of the probe.
After the probe is properly stabilized, chemically zeroed (for
measurements below 1 mg/L), and calibrated, take measurements
as follows:
1. Add the weight assembly to the probe if required (3 or 15 m
cable versions only).
2. If the sample salinity has been measured using an Electrolytic
Conductivity Meter, enter the value in setup 4. If the meter
has been moved to a different elevation or if the barometric
pressure has changed, enter the new values in setups 5 and 6.
44
ion
6 DO system:

Advertisement

Table of Contents
loading

Table of Contents