Configuring the Balance

Purpose
You can configure your Master
balance to meet individual
requirements by entering user data
and setting parameters in the
Setup menu.
The Setup menu is divided into three
functions: the application menu,
balance data and user data input.
You can also configure the display
to show specific information about
the balance (serial no., etc.).
Configuring the Balance for Use in
Legal Metrology
Set the switch as described below
to configure the following functions for
use of the balance in legal metrology:
– Display: Verification scale
interval:
;
e
lower limit of the weighing
capacity:
Min
– External calibration blocked
– MP8-interface emulation blocked
Remove the covering plate from the
back of the balance housing
Move Switch 1 in the direction of
the arrow
> Switch up: external calibration
blocked (factory setting on verified
balances)
Switch down: external calibration
accessible
> Note:
Do not move Switch 2
12

Setting the Language

Features
You can choose from 5 languages
pro
for the information display:
1 German
2 English (factory setting)
3 English with US date/time format
4 French
5 Italian
6 Spanish
Selecting the Language
Enter the corresponding number
Press M
Exit the Setup menu:
Press the oo soft key
Entering User Data
(Input)
Purpose
To display, input or change user data.
You can block access to these data
by assigning a password.
Features
You can display, input or change the
following user data:
– Workstation number for the
balance: ID (balance ID; max. 20
characters)*
– Weighing series number,
to designate a series or lot: L ID
(Lot ID; max. 20 characters)*
– Weight set number for
calibration/adjustment: W ID
(weight ID; max.14 characters)*
– Exact calibration weight value
for calibration/adjustment of the
balance, e.g. for adjustment
according to a DKD certificate (see
the section on "Calibration/
Adjustment" in the chapter entitled
"Operating the Balance")
– Time (hh.mm.ss; hh can be entered
without a preceding zero)
– Date (dd.mm.yy, or mm.dd.yy
when you select "English with US
date/time" as the language)
– Contrast/angle of the display
(enter a number from 0 to 4;
factory setting: 2)
– Password for access to the Setup
menu, which contains the Input,
Application and Balance Menu
functions (max. 8 characters)*
– ID codes (20 characters max.):
the factory-set designations
for these codes are ID1, ID2, ID3
and ID4*
The ID codes are left-justified
on the printout.
– Three calibration/adjustment
times (hh.mm; hh can be entered
without a preceding zero)
Calibration/adjustment is repeated
every day at the selected
time settings (up to 3 maximum:
for example, at 7:30, 12:00
and 18:15)
*: If the last character of user
data is a letter: conclude input by
pressing a
To delete user data: Enter a .
(decimal point) or a space and
confirm
To delete the last character
entered: Press c (see the
section on "Basic Settings" in
the chapter entitled "Operating
the Balance")