Sartorius la series Operating Instructions Manual page 59

Electronic analytical and precision balances masterpro series
Hide thumbs Also See for la series:
Table of Contents

Advertisement

Practical Example
Determining Animal Weight With Automatic Start of 20 Subweighing Operations for Averaging; Automatic Printout of the
Number of Subweighing Operations and of the Animal Weight
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Animal weighing: Animal activity: Active
Setup: App: Application 1: Animal weighing: Printout: Average and calculated values
Setup: App: Basic settings: Printout configuration: Auto print upon initialization: On, all values
Step
1. Delete previous setting if necessary
2. Prepare a container (cage)
3. Tare the balance
4. Enter number of subweighing
operations for averaging
5. Save number
6. Weigh the first animal
7. Start automatic animal weighing
The balance delays starting the
subweighing operation until
three successive subweights lie
within the range defined for
an "active" animal
Key (or instruction)
c
Place empty cage
on the balance
t
2 0
soft key
mDef
Place 1st animal in cage
soft key
Start
When this criterion is met, the
subweighing series begins
Operating the Balance
Display/Output
weight value fluctuates due to animal activity
59

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents