Sartorius la series Operating Instructions Manual page 47

Electronic analytical and precision balances masterpro series
Hide thumbs Also See for la series:
Table of Contents

Advertisement

Determination of the
Repeatability (reproTEST)
Definition
Repeatability is the ability of the
balance to display identical readouts
when it is loaded several times
with the same weight under constant
ambient conditions.
The standard deviation for a given
number of measurements is used
to quantify the repeatability.
Check the Reproducibility of the Balance
Step
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
1. If
is set:
reproTEST
and proceed with step 4.
If
Selection mode
2. Select reproTEST
3. Start reproTEST
4. Number of measurements
is displayed; 6 measurements will
now be performed
The standard deviation
is displayed
5. End reproTEST
or restart reproTEST
Purpose
The "reproTEST" function auto-
matically determines the
reproducibility of results (based on
6 individual measurements). In this
way, the balance determines one
of the most important quantities
in relation to the place of installation.
The results are displayed with the
balance's accuracy.
Key(s) (or instruction)
soft key
Cal
is set:
soft key
Cal
soft key
Select
soft key
Start
soft key
End
soft key
Start
Operating the Balance
Preparation
Turn on the balance:
Press e
> The Sartorius logo is displayed
> The balance performs a self-test
Select reproTEST in the Setup
menu: Press M
Select the Balance menu:
Press the
Menu
Select either
reproTEST
or
Selection mode
(factory setting): see "Configuring
the Balance"
Exit the Setup menu:
Press oo soft key
Display/Output
Max4200 g
0%
F
OO6
S
CAL: reproTEST
Start
soft key
d= 0.01g
100%
A
H
a
g
L
s
S
T
Wahl
47