Practical Example
Weighing in Components
Settings (changes in the factory settings required for this example):
Setup: App: Application 3: Formulation: Automatic storage: On, first value at stability
Setup: App: Application 3: Formulation: Minimum load for automatic storage: 100 digits
Setup: App: Application 3: Formulation: Evaluation mode, MR function: Final evaluation, print
Step
1. Turn on the balance and configure
the settings as indicated above
2. Delete old formulation data,
if necessary
3. Tare the balance
4. Place the empty container
on the balance (here: 180.59 g)
5. Tare the balance
6. Weigh in the first component
(here: 42.88 g)
7. Store components in the
formulation memory
Balance is tared automatically
Components are printed out
automatically
92
Key (or instruction)
e
c
t
Place load on the balance
t
Place components
in container
soft key
M+
Display/Output
--------------------
16.01.1997
Comp1 +
42.88 g
s
2
14:04