Sartorius la series Operating Instructions Manual page 88

Electronic analytical and precision balances masterpro series
Hide thumbs Also See for la series:
Table of Contents

Advertisement

Practical Example
Totalize counted pieces
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Counting
Setup: App: Application 3: Totalizing: Evaluated values: Net + calculated
Setup: App: Application 3: Totalizing: Evaluation mode, MR function: Final evaluation, display + print
Step
1. Turn on the balance and configure
the settings as indicated above
2. Delete old totalization data,
if necessary
3. Tare the balance
4. Toggle to Application 1: Counting
5. Place the displayed number of
parts on the balance (here: 10 pcs)
6. Initialize the Counting application
7. Remove the reference sample
quantity and toggle to Totalizing
8. Place a number of parts on the
balance (here: 50 pcs)
88
Key (or instruction)
e
c
t
D
Place parts to be counted
on the balance
soft key
Start
Unload the balance
D
Place parts on the balance
Display/Output
nRef
wRef
35.28900 g
10 pcs