Totalizing - Sartorius la series Operating Instructions Manual

Electronic analytical and precision balances masterpro series
Hide thumbs Also See for la series:
Table of Contents

Advertisement

Totalizing

L
Purpose
This application program acts as
a cumulative memory function.
You can use the "Totalizing"
application in combination with a
program chosen from Application 1
(e.g., counting, weighing in percent)
and one from Application 2 (check-
weighing, time-controlled functions)
as well as with the extra functions.
Available Features
– Totalization of weight values
and calculated values
– Optional configuration in
the Setup menu for simultaneous
storage of net and calculated
values
– Optional configuration in the
Setup menu for loading weight
values and calculated values
either from Application 1 (e.g.,
counting, weighing in percent)
or from Application 2 (check-
weighing, time-controlled functions)
– Totalization memory for up to
65535 values
– Simultaneous display in the text
line of transaction counter and,
e.g., the current total
– Optional configuration in
the Setup menu for having the
balance tare automatically
after a value is stored in the
totalization memory, if no preset
tare has been entered
– Manual input of the number of
individual weighing operations
(target no. of operation
nDef
Result printed and memory
cleared after printout of
nDef
– Optional configuration in the
Setup menu to add the current
weight, with display accuracy,
to the current total by pressing
the
soft key and generate
M+
a printout of the result
86
– Optional configuration in the
Setup menu for stability-dependent
storage of the measured value:
Stability range
– Optional automatic storage of
measured values
Storage of measured value is
indicated by
Uu
indicates that you can place
zz
a load on the balance.
– Minimum load threshold for
automatic storage
– Press the
soft key to delete the
M-
last value added to the totalization
memory. The transaction counter
value is reduced by one and a
printout is generated.
– Press the
soft key for inform-
MR
ation about number of transactions
and the current total. By
configuring the Setup menu, you
can define whether the information
is displayed and printed, or
only printed, and whether the
information comprises an
intermediate or final evaluation
(see the example on page 69)
– In the Info window you can
choose which value is displayed
in the text line during weighing
– Printout of the end result indepen-
dent of which program is confi-
gured for Application 1 or
Application 2. Configure the
Setup menu to define which
values are included on the
printout (Printout of individual
components)
– Press the key identified by
(soft key label) for a printout of an
intermediate evaluation after each
addition or a final evaluation
).
– If you end the totalization process
by pressing c without having
.
first pressed the
MR
for a printout, a final evaluation
is printed when you press c
– Optional configuration in the
Setup menu to clear the totaliz-
ation memory and reset the
transaction counter by pressing
c or after an evaluation is
printed out
– Totalization data and transaction
counter data are stored in non-
volatile memory
– Continue totalization after turning
the balance off and back on
Factory Settings
.
Automatic storage:
Minimum load for automatic storage:
10 digits
Source of data for auto storage:
Application 1
Evaluated values:
Evaluation mode, MR key function:
Intermediate
evaluation, print
M+/M– function, then tare:
Printout of individual components:
Balance functions:
Stability range:
1-3-4 2 increments
Tare after individual printout:
Soft Key Functions
M+
M-
MR
MR
nDef
soft key
Printout for Totalizing
The transaction or component counter
is printed in front of each measured
value (weight). When an intermediate
or final evaluation is printed out, all
results up to this point are included.
n
Total + 151.67 g
Off
Net
Off
On
No
Add weight values or
application values to the
total in the totalization
memory. The component
or transaction counter value
increases by one each time
you press this key.
Delete the last value added
to memory. The transaction
counter value decreases by
1. You cannot delete
previous values by
repeatedly pressing this key.
Print or display an inter-
mediate or final evaluation
Store the input number
of components
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents