You can have the results of a
calibration/adjustment procedure
printed out. You can configure
whether the printout is generated as
soon as the procedure is completed,
or whether a number of calibration/
adjustment procedures (up to 50)
are collected for a block printout.
Loading Stored Data:
Data for the block printout are stored
in battery-backed memory. These
data remain in memory for approx. 3
months after the equipment is dis-
connected from AC power. Make
sure to generate a printout before dis-
connecting the equipment for
a long period of time.
Block Printout of
Calibration/Adjustment Data
With the following Setup menu confi-
guration, you can store the data from
up to 50 calibration/adjustment pro-
cedures and have them printed on
request:
– isoCAL printout [1-17- ]
2 On request, from
record memory
When the memory contains 50 data
records:
– additional records are output
automatically
If at least one block printout data
record has been configured, the
following soft keys are available after
you press the
soft key:
Cal
The number of records is
Info
displayed in the text line
Print accumulated records
PrtPro
Delete accumulated
DelPro
records; records can only
be deleted after a printout
has been generated.
If a password has been
assigned in the Setup:
Input menu, you must enter
either the configured pas-
sword
or the General Password
before you can delete the
records.
For internal calibration/adjustment,
the initialization mode of the
procedure is displayed in the
Start line.
--------------------
13.05.1997
09:17
SARTORIUS
Mod.
FC6CCE-H
Ser. no.
70419914
Ver. no.
01-35-16
ID
--------------------
24.04.1997
12:03
Start:
manual
Diff. +
0.01 g
External calibration
completed
25.04.1997
12:10
Start:
isoCAL/temp
Diff. +
0.01 g
Internal adjustment
completed
Diff. +
0.00 g
25.04.1997
18:30
Start:
Adj.time
Diff. +
0.01 g
Internal adjustment
completed
Diff. +
0.00 g
26.04.1997
Start:
manual
Diff. +
0.01 g
Internal adjustment
completed
Diff. +
0.00 g
27.04.1997
11:53
Start:
Ext.cal.
W ID
Nom.
+ 2,000.00 g
Diff. +
0.01 g
External calibration
completed
Diff. +
0.00 g
--------------------
13.05.1997
09:17
Name:
--------------------
Operating the Balance
GLP header
List of Calibration/
Adjustment Procedures:
Example 1:
External calibration
Example 2:
isoCAL triggered by difference
in temperature
Example 3:
isoCAL at defined time
Example 4:
9:37
Internal calibration/adjustment
triggered manually
Example 5:
External calibration/adjustment
GLP footer
127