Internal Calibration/Adjustment
First set either
Internal cal./
or
adjustment
Selection
(factory setting) in the Setup
mode
menu (Balance menu: Balance
functions).
Inside the balance housing is a built-
in motorized calibration weight.
The internal calibration/adjustment
sequence is as follows:
Select the calibration function:
Press the
soft key and then
Cal
the
soft key
Start
> The internal calibration weight
is applied automatically
> The balance is calibrated
> If the setting
Calibrate,
then auto adjust
selected in the Balance menu,
the balance is now automatically
adjusted
> If the setting
Calibrate,
then manual adjust
selected in the Balance menu,
you can end "Internal cab/
adjustment" now; to start it, press
the
soft key without
Start
adjusting the balance (see
"Calibration and Adjustment
Sequence", next column)
> The internal calibration weight
is removed
> (ISO/GLP/GMP printout:
see page 126)
44
Calibration and
Adjustment Sequence
In the Setup menu, you can configure
the balance so that:
– calibration is always followed
automatically by adjustment
Calibrate, then auto
(factory setting) or
adjust
– you have the choice of ending
the sequence or starting
adjustment after calibration
Calibrate, then
manual adjust
If no deviation is determined in
calibration, or the deviation is within
the tolerance limits dictated by the
degree of accuracy you require, it
is not necessary to adjust the
balance. In this case, you can end
the calibration/adjustment
is
sequence after calibration. There
are 2 softkeys active at this point:
to start adjustment
Start
to end the sequence
End