MP8 Interface Emulation

Purpose
With the MP8 interface emulation
function you can connect peripheral
devices of the MP8 generation that
have separate AC power supplies
(such as the 73822... Data Control
terminal, a YFC... Flow Rate
Controller, a YDI 50 Z Data Input
dedicated keyboard, etc.) to your
pro
Master
balance.
Available Features
– The balance is only used to deter-
mine weights
– The interface communicates
exclusively in the MP8 binary
protocol.
– You can select application programs
for use with the MP8 under item 3
in the balance operating menu.
– You can set the Index 2 program
for MP8 can be selected under item
3 of the balance operating menu
– In the application menu, you
can only define parameters for the
keypad and display.
– The following parameters remain
accessible as before:
– Weighing parameters
[1-x-x]
– Extra functions
[8-x-x]
– Reset function
[9-x-x]
(see "Setup Parameters (Over-
view") in the chapter entitled
"Configuring the Balance" in the
Installation and Operating
Instructions)
– MP8 interface emulation is not
allowed when you use your
balance as a legal measuring
instrument (legal for trade). The
emulation function is de-activated
when the menu access switch is
sealed for legal metrology.
Factory Settings
(for MP8 functions)
MP8 application:
MP8: 3-1-1
Program index 2:
1 Ind. 2.1
Preparation
Turn on the balance: Press e
> The Sartorius logo is displayed
Switch to the MP8 interface:
Press M
Select balance operating menu:
Select and confirm the
menu reset function
then O soft key
Confirm
factory settings
Select and confirm
Set to MP8 [9-1-9]
q or Q soft key, repeatedly if necessary; then l soft key
Press the oo soft key
Parameter Settings for the MP8 Interface
Press M
Select the balance operating menu:
Select and confirm:
:
3 Application
1 Program selection
or
1 MP8: 3-1-1
...
or
9 MP8: 3-1-9
or
10 MP8: 3-2-1
...
or
18 MP8: 3-2-9
or
19 MP8: 3-3-1
...
27 MP8: 3-3-9
4 Program index
or
1 Ind. 2.1
or
2 Ind. 2.2
or
3 Ind. 2.3
4 Ind. 2.4
:
5 Interface
1 Baud rate
or
1 150 baud
or
2 300 baud
or
3 600 baud
or
4 1,200 baud
or
5 2,400 baud
or
6 4,800 baud
or
7 9,600 baud
2 Parity
or
1 Mark
or
2 Space
or
3 Odd
4 Even
6 Print in weighing mode
1 Manual/auto print mode
1 Manual without stability
2 Manual with stability
4 Automatic without stability
5 Automatic at stability
Store settings and exit the Setup menu: Press the oo soft key
Operating the Balance
soft key
Menu
: O soft key
soft key
Menu
;
:
2 Ind. 2
:
or
or
or
: q soft key 6 times,
:
135