Sartorius la series Operating Instructions Manual page 85

Electronic analytical and precision balances masterpro series
Hide thumbs Also See for la series:
Table of Contents

Advertisement

Practical Example
Document the amount of evaporation of a sample with defined surface, temperature and air pressure at preset intervals
of 1 minute, 30 seconds.
Settings (changes in the factory settings required for this example):
Setup: App: Application 2: Time-controlled functions
Setup: Menu: Balance functions: Taring: Without stability
Setup: Menu: Print in weighing mode: Manual/auto print mode: Manual without stability
Step
1. Turn on the balance and configure
the settings as indicated above
2. Delete stored values if necessary
3. Place container with sample
on the balance and tare
4. Enter time interval: 1 minute,
30 seconds
5. Store time interval
6. Begin documentation
(Time remaining until the
next printout is displayed
in the text line)
Printout of evaporation amount
every 1
1
/
minutes
2
7. Stop the documentation procedure
Key (or instruction)
e
c
t
1 . 3 0
soft key
Interv
soft key
Start
soft key
Stop
Operating the Balance
Display/Output
85