Practical Example
Create a New Base Data Record for Initializing the Checkweighing Program, Including: Target Value, Minimum, Maximum
Settings (changes in the factory settings required for this example):
Setup: App: Extra function (F4): Product data memory
Setup: App: Application 2: Checkweighing
Step
1. If necessary: turn on the balance
and enter the settings given above
2. In the Checkweighing application,
toggle to the input mode for target,
minimum and maximum values
3. Enter target: 170 g;
minimum: 165 g;
maximum: 180 g
4. Toggle to display of product data
(existing data records are displayed;
in this example, 3 data records have
been stored)
5. Enter a name for the new data
record (here: CHW01)
6. Store current Checkweighing
parameters as a data record
7. Confirm
8. Exit data record display
Key (or instruction)
e
soft key
Param.
see the Practical Example
for Checkweighing,
steps 5 through 9
soft key
ProDat
a
soft key,
soft key
ABCDEF
C
soft key,
soft key
GHIJKL
H
soft key,
soft key
STUVWX
W
0 1
soft key
New
soft key
Store
oo soft key