External Calibration/Adjustment* with
a User-Defined Calibration Weight
First set either
Ext. cal./
adj.;user-defined wt.
or
(factory
Selection mode
setting) in the Setup menu: Balance
functions.
You can define a weight for
calibration/adjustment. External
calibration/adjustment must
be performed with weights that are
traceable to a national standard
Define the Calibration Weight
Step
–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
1. Select Setup
2. Select Input
3. Select calibration/adjustment
weight
4. Enter calibration weight
(e.g., 4000.00 g)
and store
6. Exit the Setup menu
* for balances of accuracy class K, only "external adjustment" is possible
and that have error limits which are
at least 1/3 of the required
tolerance of the display accuracy.
The defined weight must equal
at least 10% of the maximum
balance capacity.
See page 43 for the external
calibration/adjustment sequence.
For this example, select external
calibration/adjustment with a
user-defined weight.
The balance has a factory-set weight
value (see "Specifications").
Key(s) (or instruction)
M
soft key
Input
q soft key 3 times
4 0 0 0 . 0
0
l soft key
oo soft key
Operating the Balance
To reset a user-defined calibration
weight to the original factory setting:
$ Enter the factory-defined value
manually (see "Specifications")
Display/Output
3000.00 = last setting selected
Max4200 g
0%
F
000
S
Cal
d= 0.01g
100%
A
H
a
g
L
s
S
T
45