Practical Example 2: Animal weighing with statistics
Determine the weights of 7 mice; generate and print a statistical evaluation.
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Animal weighing: Start: Automatic
Setup: App: Application 1: Animal weighing: Printout: Off
Setup: App: Application 3: Statistics: Automatic storage: On, first value at stability
Setup: App: Application 3: Statistics: Minimum load for automatic storage: 100 digits
Setup: App: Application 3: Statistics: Evaluated value: Calculated
Setup: App: Application 3: Statistics: Evaluation mode, MR function: Intermediate evaluation, display+print
Setup: App: Extra function(F4): Man. store in app. 3 memory (M+)
Step
1. Prepare a container (cage)
2. Tare the balance
3. Enter number of subweighing
operations for averaging
4. Save number
5. Weigh the first animal
6. Start automatic animal weighing
116
Key (or instruction)
Place empty cage
on the balance
t
2 0
soft key
mDef
Place 1st animal in cage
soft key
Start
Display/Output
weight value fluctuates due to animal activity