2 Environmental Requirements/Limitations

2.1 Room Climate

Avoid extremes in temperatures
Innova system room climate requirements – relative humidity and temperature (individual
products or components are classed by their installation area)
Table 3-1:
INSTALLATION
ROOM OF PRODUCT
OR COMPONENT
MIN
Examination room
30%
Technical room (See
30%
Note (1))
Technical room with
30%
Fluoro UPS optional
Control room
30%
Innova Digital Detector
See Note (2)
NOTE:
(1): The target temperature (best recommended) is 18 C (64 F).
NOTE:
(2): The detector should be stored at 10 to 40 C (50 to 104 F) and less than or equal to
90% RH in the plastic wrapped shipping box. (This should include two bags of desiccant as
well). The lowest temperature (e.g. 10 ºC (50 F)) and humidity is preferable. If they are to be
stored outside of their shipping box or in the inner shipping box without plastic wrapping they
should be stored at 20 C (68 F) or less and 30% RH or less. In terms of transportation,
do not expose to temperatures below –20 ºC (–4 F) in its shipping box for more than 15
hours. The detector will reach the ambient temperature after 20 to 25 hours. The detector
should not be allowed to reach temperatures less than –10 ºC (14 F) or irreparable damage
to the detectors scintillator will occur. Care must be taken when removing a detector from a
shipping box. If the detector has been subject to cold temperatures for an extended period
the detector in the box should be allowed to sit in the plastic wrapped box to reach room
temperature. This will prevent condensation from occurring. Condensation on the detector
can cause irreparable damage to the electronics. Storage 10 to 40 ºC (50 to 104 F); 10 to 90
% RH, 250 day storage transportation –20 to +60 ºC (–4 to 140 F) and 10 to 80% RH. The
Detector chiller HEAT DRY 1 is shipped within GEMS packaging.
NOTE:
(3): In use temperature limits specify the range where the system shall work. Operating
outside these limits could occur severe performance and reliability issues.
Recommended temperature limits specify the range where it is recommended to adjust air
conditioning control in order to warranty current operations inside the in use range.
46
RELATIVE HUMIDITY (NON-
CONDENSING)
IN-USE
STORAGE
MAX
MIN
70%
10%
75%
10%
75%
10%
75%
10%
IN-USE (See Note
(3))
MAX
MIN
MAX
+15 C
+32 C
70%
+59 F
+90 F
+10 C
+35 C
70%
+50 F
+95 F
+20 C
+25 C
70%
+68 F
+77 F
+10 C
+35 C
70%
+50 F
+95 F
2 Environmental Requirements/Limitations
Direction 5314528-2-1EN, Revision 1
TEMPERATURE
RECOMMENDED
(See Note (3))
MIN
MAX
MIN
Design for Patient/
-10 C
Operator Comfort
+14 F
+13 C
+25 C
-14 C
+55 F
+77 F
+8 F
+20 C
+25 C
-5 C
+68 F
+77 F
+23 F
+20 C
+25 C
-10 C
+68 F
+77 F
+14 F
STORAGE /
TRANSPORT
MAX
+50 C
+122 F
+50 C
+122 F
+50 C
+122 F
+50 C
+122 F