Sample period out of bounds: <bounds>
Cause: The external timing sample period must be between 10 ns and 50 ms (in a
1/2/5 sequence).
Action: Re-enter the command with the sample period between the bounds shown.
2030
Negated patterns not allowed in timing
Cause: This error occurs when you attempt to specify a "not equals" expression
when defining the external timing trigger. You can only specify labels which equal
patterns (of 1's, 0's, or X's).
Action: Do not attempt to specify negated timing patterns.
2031
Invalid trigger duration: <duration>
Cause: This error occurs when you attempt to specify an external timing trigger
duration which is in the valid range but is not a multiple of 10 ns.
Action: Re-enter the command with the trigger duration as a multiple of 10 ns.
2032
Trigger duration out of bounds: <bounds>
Cause: This error occurs when you attempt to specify an external timing trigger
duration outside the valid range. A "greater than" duration must fall within the
range of 30 ns to 10 ms (and must be a multiple of 10 ns). A "less than" duration
must fall within the range 40 ns to 10ms (and must be a multiple of 10 ns).
Action: Re-enter the command with the trigger duration within the bounds shown.
2042
Trigger delay out of bounds: <bounds>
Cause: This error occurs when you attempt to specify an external timing trigger
delay outside the valid range. The external timing trigger delay must be between 0
and 10 ms (in 10 ns increments).
Action: Re-enter the command with the trigger delay within the bounds shown.
Chapter 13: Error Messages
Terminal Interface Messages
525