4
Basic Operation

Preparing the LC Modules

Preparing the LC Modules

While conditioning or equilibrating the column, it is possible to tune the Triple
Quad LC/MS.

Switch LC stream to Waste

When not acquiring data, switch the direction of the LC stream away from the
MS ion source and to waste.
When the is LC connected to a VWD or DAD, monitor the fluctuations of the
VWD or DAD real-time chromatogram before a run.
a Right-click the TQ device in the Instrument Status window.
b Select LC > Waste.
4 Purge the LC pump.
• Follow the specific user guide directions for purging the installed pump
5 Set up to condition or equilibrate the column.
a In the Method Editor window, click Apply in the toolbar to download the LC
Agilent Triple Quadrupole LC/MS System User Guide
model.
parameters to the LC.
70