Testing Performance

To re-assign pods reserved for time tag storage

Conclude the State Mode Tests

66
From the main menu, select Setup→Pod Assignment... or
1
Setup→My Logic Analyzer-1→Pod Assignment...
In the Pod Assignment dialog, select pod pair A1/A2 to be
2
reserved for time tag storage.
Click OK.
3
Reconfigure the markers to search for 131071 occurrences
4
instead of 131072 (see
page 52).
After re- assigning pod pairs, you can test pods that were
previously reserved for time tag storage.
Note that you can reserve Pod 1 and Pod 2 on the master
card for time tag storage and still use the CLK input on
Pod 1.
Do the following steps to properly shut down the logic
analyzer session after completing the state mode tests.
1
End the test.
a
In the Agilent Logic Analyzer application, select the
[X] in the upper right corner to close the window. At
"Configure the markers"