Setting Baud Rate; Connecting The N-7000 Interface; Baud Rate Switch Settings - Nellcor N-200 Operator's Manual

Hide thumbs Also See for N-200:
Table of Contents

Advertisement

Setting Baud Rate

Identify the baud rate required by the device that is to be
connected, and set the N-200 baud rate using the rear-panel
baud rate switches (switch sections 7 and 8). Refer to Table 6.

CONNECTING THE N-7000 INTERFACE

Note:
The NELLCOR N-7000 is not available outside the
United States.
The N-7000 provides RS-232 or RS-485 interface capabilities to
the N-200. Data from the N-200 is transmitted through a fiber-
optic cable to the N-7000. The N-7000 outputs data via a
standard 25-pin Dconnector.
Connect the N-7000 to the N-200 as follows:
1.
Set the N-200 BAUD RATE switches to 2400 (switch 7
down; 8 up).
2.
Set the N-200 RS-232 FORMAT switches to the N-7000/
N-9000 format (switches 3–5 down).
3.
Connect the gray end of the fiber-optic cable provided with
the N-7000 to the DATA connector on the N-200
powerbase. Connect the blue end to the unlabeled data
input connector on the side of the N-7000. Push each
connector in until it snaps into place. Route the fiber-optic
cable without bending it; poor transmission may occur if
the cable is kinked.
Table 6: Baud Rate Switch Settings
Baud Rate
1200
2400
9600
19200
Connecting to Other Instruments
Switch Sections
7
Down
Down (off)
Up (on)
Up (on)
8
Down
Up
Down
Up
65