2.
SPECIFICATIONS AND AGENCY
APPROVALS
2.1.
Specifications
PARAMETER
Ranges
Measurement Units
Zero Noise
Span Noise
Lower Detectable Limit
Accuracy
Response Time (95%)
Sample Flow Rate
Temperature Range
Humidity Range
Dimensions (H x W x D)
Weight
Power
Environmental Conditions
Analog Output, Voltage Mode
05509G DCN6806
DESCRIPTION
User selectable to any full scale range from:
1 PPM to 500 PPM
1000 PPB to 10,000 PPB
3
2000 µg/m
to 10,000 µg/m
100 PPHM to 10,000 PPHM (optional configuration)
Low conc scrubber configuration: 0 to 20 PPM
High conc scrubber configuration: >20 PPM
3
PPM, PPB, µg/m
, PPHM (optional configuration)
< .0015 PPM (rms)
< .5% of reading (rms) (above 0.1 PPM)
< .003 PPM (rms)
+/- 1% of full scale range
<30 sec
(<60 sec with the Sample Conditioner option)
0.8 L/min
5-45 o C
10-90% RH, Non-Condensing
5.22" x 19.0" x 15.3" (133 mm x 483 mm x 388 mm)
– Rack Mount Version
16.01" x 15.6" x 6.90" (407mm x 396mm x 175mm)
– with NEMA 4X enclosure
12.6 lb. (5.73 kg) - Rack Mount Version
15.2 lb. (6.91 kg) – with NEMA 4X enclosure
100-240 V~, 47-63 Hz, 74 W
Installation Category (Overvoltage Category) II
Pollution Degree 2
For Indoor Use Only
Maximum Operating Altitude 2000 meters
0 - 5V
Teledyne API Model 465L O
3
3
Monitor Manual
13