To remove the Fan assembly

7000 Series Oscilloscopes Service Guide
1 Insure that fan power cable has been disconnected from
Acquisition board.
2 Carefully slide fan assembly (fan and fan mount) to the
right, then lift away from rear deck. Note, fan mount is
soft and can be damaged by sharp sheet metal edges.
Take care that the fan power cable is not damaged when
pulling across sheet metal edges.
Removing fan assembly
Figure 35
Replacing Assemblies