Step 2. Remove the Front Panel Assembly

Refer to
Figure 4
for this procedure.
1. With a 5/16-inch wrench, remove the six semi-rigid jumpers (item
2. With a T-10 TORX driver, remove the eight screws (item
Before removing the front panel from the analyzer, lift and support the front of
CAUTION
the analyzer chassis.
3. Slide the front panel over the test port and other front panel connectors.
4. Disconnect the front panel interface ribbon cable (item
from the analyzer.
Figure 4
Front Panel Assembly Removal
Installation Note N5230-90022
x
) from the front panel.
y
) from the sides of the frame.
z
). The front panel is now free