Agilent Technologies 11644A Series User's And Service Manual

Waveguide calibration kits, serial number prefix 3032a
Table of Contents

Advertisement

Quick Links

User's and Service Guide
Agilent Technologies 11644A
X, P, and K
Waveguide Calibration Kits
This manual applies to 11644A series calibration kits with
serial number prefix 3032A.
Agilent Part Number: 11644-90371
Printed in USA
Print Date: June 2006
Supersedes: October 2002
© Copyright 1994, 1996, 1997, 2001–2002, 2006 Agilent Technologies, Inc. All rights reserved.

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 11644A Series

  • Page 1 This manual applies to 11644A series calibration kits with serial number prefix 3032A. Agilent Part Number: 11644-90371 Printed in USA Print Date: June 2006 Supersedes: October 2002 © Copyright 1994, 1996, 1997, 2001–2002, 2006 Agilent Technologies, Inc. All rights reserved.
  • Page 2 THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL. Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent products. For any assistance, contact Agilent Technologies. Refer to “Contacting Agilent” on page 6-4.
  • Page 3: Table Of Contents

    11644A Series Kits with Serial Prefix 3012A ........
  • Page 4 Contents Tightening a Flange Connection ..........4-8 Inspecting a Flange Connection .
  • Page 5: General Information

    General Information X, P, and K 11644A...
  • Page 6: Calibration Kit Overview

    General Information Calibration Kit Overview Calibration Kit Overview This waveguide calibration kit is used to calibrate network analyzer systems (such as the Agilent 8510, 872x, or PNA series). With the calibration data properly loaded in the network analyzer and a measurement calibration completed, systematic errors are minimized.
  • Page 7: Equipment Required But Not Supplied

    Calibration Kit History This manual applies to any Agilent 11644A series (X, P, or K) waveguide calibration kits whose serial number prefix is 3032A and above. If your calibration kit has a different serial number prefix, refer to the next section for information on how this manual applies.
  • Page 8: Recording The Device Serial Numbers

    General Information Recording the Device Serial Numbers Recording the Device Serial Numbers In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in the appropriate table.
  • Page 9: Preventive Maintenance

    General Information Preventive Maintenance Preventive Maintenance The best techniques for maintaining the integrity of the devices in the kit include: • routine visual inspection • routine cleaning • proper gaging • proper connection techniques All of these are described in Chapter 4.
  • Page 10 General Information Preventive Maintenance X, P, and K 11644A...
  • Page 11 Specifications X, P, and K 11644A...
  • Page 12: Specifications Environmental Requirements

    Specifications Environmental Requirements Environmental Requirements Table 2-1 Environmental Requirements Parameter Limits Temperature +20 °C to +26 °C Operating −40 °C to +75 °C Storage Error corrected range ±1 °C of measurement calibration temperature Altitude Operating < 4,500 meters (≈15,000 feet) Storage <...
  • Page 13: Mechanical Characteristics

    Specifications Mechanical Characteristics Mechanical Characteristics Table 2-2 Table 2-3 list the typical characteristics of the adapters and ¼ wavelength shims in your kit. These are not specifications, but are included as additional information. Adapters Table 2-2 Mechanical Characteristics of Adapters Frequency band Insertion Center Conductor...
  • Page 14: Shorts

    Specifications Mechanical Characteristics ¼ Table 2-3 Mechanical Characteristics of Wavelength Shims Part Number Waveguide Thickness Band (mm) (mm) (mm) (mm) 11664-20021 WR90 22.86 ±0.01 10.16 ±0.01 0.13 max 9.78 ±0.02 11664-20020 WR62 15.80 ±0.01 7.90 ±0.01 6.50 ±0.02 11664-20019 WR42 10.671 ±0.008 4.318 ±0.008 4.5 ±0.02...
  • Page 15: Electrical Specifications

    Specifications Electrical Specifications Electrical Specifications Table 2-4 through Table 2-6 list the electrical specifications of the terminations, ¼ wavelength shims, and standard sections in your kit. Table 2-4 Electrical Specifications of Terminations Device Frequency Termination Specification return loss ≥ 42 dB X11644A WR-90 8.2 to 12.4 GHz return loss ≥...
  • Page 16: Certification

    Technical Papers hyperlink. Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities...
  • Page 17: User Information

    User Information X, P, and K 11644A...
  • Page 18: Calibration Devices And Their Use

    User Information Calibration Devices and Their Use Calibration Devices and Their Use The P, K, and X11644A waveguide calibration kits contain termination load, standard sections, shim, and short. For measurement convenience, these kits contain either 7 mm coax-to-waveguide adapters for P-band and X-band, or 3.5 mm coax-to-waveguide adapters for K-band. The adapters are intended for adapting coaxial test sets to waveguide.
  • Page 19: Measurement Applications

    User Information Measurement Applications Measurement Applications For your convenience, two different lengths of screws are provided in this kit. NOTE While you can use the long screws for any connection, the shorter screws provide a faster connection for two-flange connections. Isolation In most cases, select the softkey on your network analyzer.
  • Page 20: Load

    User Information Measurement Applications Load Connect the termination to the appropriate test port, as shown inFigure 3-3. Figure 3-3 Test Port and Termination Offset Load Connect the shim and the termination to the appropriate test port, as shown inFigure 3-4. Figure 3-4 Connect the Shim, Termination, and Test Port X, P, and K 11644A...
  • Page 21: Offset Short

    User Information Measurement Applications Offset Short Connect the shim and the short to the appropriate test port, as shown inFigure 3-5. Figure 3-5 Connect the Shim, Short, and Test Port Reflect Connect the short to the appropriate test port, as shown inFigure 3-6.
  • Page 22: Thru

    User Information Measurement Applications Thru No device is required for this. Connect port 1 to port 2, as shown inFigure 3-7. Figure 3-7 Port 1 and Port 2 ¼ Wavelength Load “Offset Load”. ¼ Wavelength Short “Offset Short”. X, P, and K 11644A...
  • Page 23: Changing The ¼ Wavelength Shim Calibration Definition

    User Information Changing the ¼ Wavelength Shim Calibration Definition Changing the ¼ Wavelength Shim Calibration Definition The calibration kit definition data provided with this kit has a nominal value for the ¼ wavelength shim offset delay. You may use the nominal value provided, or measure the exact thickness of the shim and use that value to calculate its exact offset delay.
  • Page 24 User Information Changing the ¼ Wavelength Shim Calibration Definition X, P, and K 11644A...
  • Page 25: Use, Maintenance, And Care Of The Devices

    Use, Maintenance, and Care of the Devices X, P, and K 11644A...
  • Page 26: Electrostatic Discharge

    Use, Maintenance, and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged.
  • Page 27: Visual Inspection

    Use, Maintenance, and Care of the Devices Visual Inspection Figure 4-1 ESD Protection Setup Visual Inspection Visual inspection and, if necessary, cleaning should be done every time a connection is made. Inspect mating surfaces for dirt, dust, foreign particles, or scratches, which can degrade device performance.
  • Page 28 Use, Maintenance, and Care of the Devices Cleaning the Mating Plane Surfaces These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information. WARNING Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container.
  • Page 29: Connections

    Use, Maintenance, and Care of the Devices Connections Connections Good connections require a skilled operator. Slight errors in operator technique can have a significant effect on measurements and measurement uncertainties. The most common cause of measurement error is poor connections. The following procedures illustrate how to make good connections.
  • Page 30: Aligning Two Precision Flanges

    Use, Maintenance, and Care of the Devices Connections Aligning Two Precision Flanges 1. Place the slip pins in the top and bottom holes of one flange, as shown in Figure 4-3. Figure 4-3 Inserting Slip Pins 2. Using the pins as guides, carefully align the flanges and insert two screws in the diagonal corner holes, as shown in Figure 4-4.
  • Page 31: Aligning A Precision And A Non-Precision Flange

    Use, Maintenance, and Care of the Devices Connections Aligning a Precision and a Non-Precision Flange 1. Place an alignment pin (with head) in the corner hole of one flange. Place a second alignment pin in the diagonal corner hole of the second flange, as shown in Figure 4-5.
  • Page 32: Tightening A Flange Connection

    Use, Maintenance, and Care of the Devices Connections Tightening a Flange Connection NOTE The best connection has symmetrical pressure applied as you gradually tighten the screws. 1. In an “X” pattern (for equal compression), tighten all four screws using a hex ball driver. Do not over-tighten.
  • Page 33: Handling And Storage

    Use, Maintenance, and Care of the Devices Handling and Storage Handling and Storage • Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use. • Never store devices loose in a box, or in a desk or bench drawer. This is the most common cause of device damage during storage.
  • Page 34 Use, Maintenance, and Care of the Devices Handling and Storage 4-10 X, P, and K 11644A...
  • Page 35: Performance Verification

    Performance Verification X, P, and K 11644A...
  • Page 36: Introduction

    These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Agilent Technologies. X, P, and K 11644A...
  • Page 37: Recertification

    A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. Refer to “Contacting Agilent” on page 6-4 for a list of offices.
  • Page 38: Performance Test

    Performance Verification Performance Test Performance Test Termination Return Loss Measurement For your convenience the following procedures have been included in this kit. Use this test, as needed, to check the performance of the terminations in this kit. Figure 5-1 Return Loss Test Setup 1.
  • Page 39: In Case Of Failure

    Performance Verification Performance Test 4. Set the averaging factor to 512. 5. At the adapter test port, perform a 2–port TRL calibration. 6. Turn on the calibration. 7. Connect the termination you wish to test to port 1. 8. Measure the return loss (S ) of the load.
  • Page 40: System Operation Checks

    Performance Verification System Operation Checks System Operation Checks Use the following procedures and the standard section in this kit to verify the operation of your calibrated system. NOTE If you are using a different analyzer, refer to its documentation for specific measurement instructions.
  • Page 41 Performance Verification System Operation Checks 7. To set the display: a. Press b. Press RESPONSE [SCALE] [.1] [x1] c. Select [REF POSN] [10] [x1] d. Select [REF VALUE] [0] [x1] e. Press MEASUREMENT [RESTART] 8. After one complete measurement sweep, the displayed trace should look similar to the typical trace shown in Figure 5-3.
  • Page 42: The S11 And S22 Of A Thru Measurement

    Performance Verification System Operation Checks The S and S of a Thru Measurement 1. Verify that the calibration is still on. 2. Verify that the averaging factor is set to 512. 3. Connect the standard section to ports 1 and 2 to form a thru connection. 4.
  • Page 43: The S21 And S12 Of A Thru Measurement

    Performance Verification System Operation Checks The S and S of a Thru Measurement 1. Verify that the calibration is still on. 2. Verify that the averaging factor is set to 512. 3. Connect the standard section to ports 1 and 2 to form a thru connection. 4.
  • Page 44: In Case Of Failure

    Performance Verification System Operation Checks In Case of Failure If a system check fails, recalibrate the system and repeat the entire procedure (all three measurements). Remember that poor connections are the most common cause of measurement errors. Also, both the hardware state and the instrument state must be correct.
  • Page 45 Performance Verification System Operation Checks Table 5-1 Using a TRL Calibration X11644A WR-90 (X-band) Test Measured Acceptable Pass Value Value (Yes/No) ≤ 0.15 dB , shorted ≤ 0.15 dB , shorted ≤ −45 dB , thru ≤ −45 dB , thru ≤...
  • Page 46: 872X System Operation Check

    Performance Verification System Operation Checks 872x System Operation Check 1. Short the standard section and perform an S and an S measurement to verify that the ripple on the reflection of the shorted waveguide section does not exceed the limits listed in the tables on page 5-15.
  • Page 47: The S11 And S22 Of A Thru Measurement

    Performance Verification System Operation Checks 9. Record the maximum peak-to-peak value in the appropriate table on page 5-15. 10. Turn off the delta marker mode. 11. Repeat this procedure from step 6, and make an S measurement on the standard section.
  • Page 48: In Case Of Failure

    Performance Verification System Operation Checks 7. Record the maximum peak-to-peak value in the appropriate table on page 5-15. 8. Repeat this procedure from step 3, and make an S measurement. Record the measured value in the appropriate table on page 5-15.
  • Page 49 Performance Verification System Operation Checks Table 5-4 Using a Full 2-Port Calibration X11644A WR-90 (X-band) Test Measured Acceptable Pass Value Value (Yes/No) ≤ 0.15 dB , shorted ≤ 0.15 dB , shorted ≤ −40 dB , thru ≤ −40 dB , thru ≤...
  • Page 50 Performance Verification System Operation Checks 5-16 X, P, and K 11644A...
  • Page 51: Troubleshooting

    Troubleshooting X, P, and K 11644A...
  • Page 52: Troubleshooting Process

    Troubleshooting Troubleshooting Process Troubleshooting Process This manual contains limited information about network analyzer system operation. For complete information, refer to the instrument documentation. If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 6-1.
  • Page 53: Returning A Kit Or Device To Agilent

    Troubleshooting Returning a Kit or Device to Agilent Returning a Kit or Device to Agilent If your kit or device requires service, contact Agilent Technologies for information on where to send it. Refer to “Contacting Agilent” on page 6-4. Include a service tag (located near the end of this manual) on which you provide the following information: •...
  • Page 54: Contacting Agilent

    Troubleshooting Contacting Agilent Contacting Agilent Assistance with test and measurement needs and information on finding a local Agilent office are available on the Web at: http://www.agilent.com/find/assist If you do not have access to the Internet, please contact your Agilent field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number.
  • Page 55: Replaceable Parts

    Replaceable Parts X, P, and K 11644A...
  • Page 56 Replaceable Parts X, P, and K 11644A...
  • Page 57: Introduction

    To order a listed part, note the description, part number, and the quantity desired. Telephone or send your order to Agilent Technologies. Refer to “Contacting Agilent” on page 6-4 for further information.
  • Page 58 Replaceable Parts Introduction Table 7-1 Replaceable Parts for the X11644A WR-90 Description Agilent Per Kit Part Number Calibration Devices Termination 00910-60003 Standard section 00896-60008 Short 11644-20018 ¼ Wavelength shim 11644-20021 7 mm coax-to-waveguide adapter X281C Option 006 Hardware Alignment pin 11644-20024 Slip pin 11644-20025...
  • Page 59 Replaceable Parts Introduction Figure 7-1 X-Band Component Identification Sheet X, P, and K 11644A...
  • Page 60 Replaceable Parts Introduction Table 7-2 Replaceable Parts for the P11644A WR-62 Agilent Description Per Kit Part Number Calibration Devices Termination 00910-60002 Standard section 00896-60007 Short 11644-20017 ¼ Wavelength shim 11644-20020 7 mm coax-to-waveguide adapter P281C Option 006 Hardware Alignment pin 11644-20023 Slip pin 11644-20025...
  • Page 61 Replaceable Parts Introduction Figure 7-2 P-Band Component Identification Sheet X, P, and K 11644A...
  • Page 62 Replaceable Parts Introduction Table 7-3 Replaceable Parts for the K11644A WR-42 Description Agilent Per Kit Part Number Calibration Devices Termination 00910-60001 Standard section 00896-60006 Short 11644-20016 ¼ Wavelength shim 11644-20019 3.5 mm coax-to-waveguide adapter (m) 00281-60001 3.5 mm coax-to-waveguide adapter (f) K281C Option 006 Hardware Alignment pin...
  • Page 63 Replaceable Parts Introduction Figure 7-3 K-Band Component Identification Sheet X, P, and K 11644A...
  • Page 64 Replaceable Parts Introduction Table 7-4 Items Not Included in the Calibration Kit Description Agilent Part Number ESD Protection Devices Grounding wrist strap 9300-1367 5 ft grounding cord for wrist strap 9300-0980 2 ft by 4 ft conductive table mat with 15 ft grounding 9300-0797 wire ESD heel strap...
  • Page 65 Standard Definitions X, P, and K 11644A...
  • Page 66: Standard Definitions Version Changes

    Standard Definitions Version Changes Version Changes Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Agilent’s Calibration Kit Definitions Web page at http://na.tm.agilent.com/pna/caldefs/stddefs.html.
  • Page 67: Standard Class Assignments

    Standard Definitions Standard Class Assignments Standard Class Assignments Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration. A class or group of classes corresponds to the systematic errors to be removed from the measured network analyzer response. Tables A-1 list the classes of the devices in the kit for the 8510 and PNA series respectively.
  • Page 68 Standard Definitions Standard Class Assignments Table A-1 Standard Class Assignments for the 8510 Network Analyzer Disk File Name: CK_WR-90, Calibration Kit Label: WR-90 A.1, or or CK_WR-62, or CK_WR-42 WR-62 A.1, or WR-42 A.1 Class Standard Class Label Short Offset Loads Short Offset...
  • Page 69 Standard Definitions Standard Class Assignments Table A-2 Standard Class Assignments for the PNA Series Network Analyzer SOLT Class TRL Class Assignments Assignments Calibration Kit Name: Calibration Kit Name: X11644A, P11644A, X11644A, P11644A, K11644A K11644A Class Class TRL thru TRL reflect 1, 2 7, 2 line/match...
  • Page 70: Blank Forms

    Standard Definitions Standard Class Assignments Blank Forms The standard class assignments may be changed to meet your specific requirements. Tables A-3 are provided to record the modified standard class assignments. Table A-3 Blank Form for the 8510 Network Analyzer Disk File Name: _____________ Calibration Kit Label: _____________ Class Standard...
  • Page 71 Standard Definitions Standard Class Assignments Table A-4 Blank Form for the PNA Series Network Analyzer SOLT Class TRL Class Assignments Assignments Calibration Kit Name: Calibration Kit Name: ____________ ____________ Class Class TRL thru TRL reflect line/match Forward Isolation transmission Cal Ref Z0 = Line Z0 Isolation Transport Ref Plane = Thru Standard...
  • Page 72: Nominal Standard Definitions

    Standard Definitions Nominal Standard Definitions Nominal Standard Definitions Standard definitions provide the definitions needed to mathematically model the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. The nominal values of these definitions are theoretically derived from the physical dimensions and material of each calibration standard, or from actual measured response.
  • Page 73 Standard Definitions Nominal Standard Definitions Table A-5 Standard Definitions for the 8510 with X-band (WR-90) = 1 Ω Calibration Kit Label: WR-90 A.1 System Z Disk File Name: CK_WR-90 Frequency Offset Standard in GHz 6.555 13.111 Short Short λ/4 Offset Short/ 32.633 6.555 13.111...
  • Page 74 Standard Definitions Nominal Standard Definitions Table A-6 Standard Definitions for the 8510 with P-band (WR-62) = 1 Ω Calibration Kit Label: WR-62 A.1 System Z Disk File Name: CK_WR-62 Frequency Offset Standard in GHz 9.485 18.97 Short Short λ/4 Offset Short/ 21.689 9.485 18.97...
  • Page 75 Standard Definitions Nominal Standard Definitions Table A-7 Standard Definitions for the 8510 with K-band (WR-42) = 1 Ω Calibration Kit Label: WR-42 A.1 System Z Disk File Name: CK_WR-42 Frequency Offset Standard in GHz 14.047 28.094 Short Short λ/4 Offset Short/ 15.015 14.047 28.094...
  • Page 76 Standard Definitions Nominal Standard Definitions Table A-8 Standard Definitions for the PNA Series with X-band (WR-90) = 1 Ω Calibration Kit Label: X-band System Z Frequency Offset Standard in GHz Offset Offset 6.555 13.111 Offset load load Load Fixed 6.555 13.111 Fixed Load Short...
  • Page 77 Standard Definitions Nominal Standard Definitions Table A-9 Standard Definitions for the PNA Series with P-Band (WR-62) = 1 Ω Calibration Kit Label: P-band System Z Frequency Offset Standard in GHz Offset Offset 9.485 18.970 Offset load load Load Fixed 9.485 18.970 Fixed Load Short...
  • Page 78 Standard Definitions Nominal Standard Definitions Table A-10 Standard Definitions for the PNA Series with K-Band (WR-42) = 1 Ω Calibration Kit Label: K-band System Z Frequency Offset Standard in GHz Offset Offset 14.047 28.094 Offset load load Load Fixed 14.047 28.094 Fixed Load Short 14.047 28.094...
  • Page 79: Blank Forms

    Standard Definitions Nominal Standard Definitions Blank Forms The standard definitions may be changed to meet your specific requirements. Tables A-11 A-12 are provided to record the modified standard definitions. Table A-11 Blank Form for the 8510 Network Analyzer Calibration Kit Label: ____________________________ System Z = _________ ____________...
  • Page 80 Standard Definitions Nominal Standard Definitions Table A-12 Blank Form for the PNA Series Network Analyzer Calibration Kit Label: K-band __________________ System Z = ____________ Frequency Offset Standard in GHz a. Ensure system Z of network analyzer is set to this value. b.
  • Page 81 7-10 calibration Agilent Technologies performance cleaning contacting how Agilent verifies compatible analyzers Agilent Technologies application verifying connecting note case, storage handling alcohol, isopropyl certificate of calibration maintenance part number 7-10 certification, device specifications...
  • Page 82 Index return loss specification standard sections handling National Institute of Standards terminations heel strap, part number 7-10 and Technology (NIST) history, serial numbers humidity specifications network analyzers, compatible serial numbers nitrogen, for cleaning devices nominal standard definitions history impedance, system non-precision flanges, aligning recording incoming inspection...
  • Page 83 Index National Institute of Standards part number and Technology (NIST) wrist strap for ESD protection static electricity part number 7-10 storage storage case, part number strap heel and wrist, part number 7-10 supplies, cleaning part number 7-10 system impedance table mat for ESD protection part number 7-10...

This manual is also suitable for:

11644x11644p11644k

Table of Contents