Page 2
THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL. Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent products. For any assistance, contact Agilent Technologies. Refer to “Contacting Agilent” on page 6-4.
General Information Calibration Kit Overview Calibration Kit Overview This waveguide calibration kit is used to calibrate network analyzer systems (such as the Agilent 8510, 872x, or PNA series). With the calibration data properly loaded in the network analyzer and a measurement calibration completed, systematic errors are minimized.
Calibration Kit History This manual applies to any Agilent 11644A series (X, P, or K) waveguide calibration kits whose serial number prefix is 3032A and above. If your calibration kit has a different serial number prefix, refer to the next section for information on how this manual applies.
General Information Recording the Device Serial Numbers Recording the Device Serial Numbers In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in the appropriate table.
General Information Preventive Maintenance Preventive Maintenance The best techniques for maintaining the integrity of the devices in the kit include: • routine visual inspection • routine cleaning • proper gaging • proper connection techniques All of these are described in Chapter 4.
Page 10
General Information Preventive Maintenance X, P, and K 11644A...
Specifications Environmental Requirements Environmental Requirements Table 2-1 Environmental Requirements Parameter Limits Temperature +20 °C to +26 °C Operating −40 °C to +75 °C Storage Error corrected range ±1 °C of measurement calibration temperature Altitude Operating < 4,500 meters (≈15,000 feet) Storage <...
Specifications Mechanical Characteristics Mechanical Characteristics Table 2-2 Table 2-3 list the typical characteristics of the adapters and ¼ wavelength shims in your kit. These are not specifications, but are included as additional information. Adapters Table 2-2 Mechanical Characteristics of Adapters Frequency band Insertion Center Conductor...
Specifications Electrical Specifications Electrical Specifications Table 2-4 through Table 2-6 list the electrical specifications of the terminations, ¼ wavelength shims, and standard sections in your kit. Table 2-4 Electrical Specifications of Terminations Device Frequency Termination Specification return loss ≥ 42 dB X11644A WR-90 8.2 to 12.4 GHz return loss ≥...
Technical Papers hyperlink. Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities...
User Information Calibration Devices and Their Use Calibration Devices and Their Use The P, K, and X11644A waveguide calibration kits contain termination load, standard sections, shim, and short. For measurement convenience, these kits contain either 7 mm coax-to-waveguide adapters for P-band and X-band, or 3.5 mm coax-to-waveguide adapters for K-band. The adapters are intended for adapting coaxial test sets to waveguide.
User Information Measurement Applications Measurement Applications For your convenience, two different lengths of screws are provided in this kit. NOTE While you can use the long screws for any connection, the shorter screws provide a faster connection for two-flange connections. Isolation In most cases, select the softkey on your network analyzer.
User Information Measurement Applications Load Connect the termination to the appropriate test port, as shown inFigure 3-3. Figure 3-3 Test Port and Termination Offset Load Connect the shim and the termination to the appropriate test port, as shown inFigure 3-4. Figure 3-4 Connect the Shim, Termination, and Test Port X, P, and K 11644A...
User Information Measurement Applications Offset Short Connect the shim and the short to the appropriate test port, as shown inFigure 3-5. Figure 3-5 Connect the Shim, Short, and Test Port Reflect Connect the short to the appropriate test port, as shown inFigure 3-6.
User Information Measurement Applications Thru No device is required for this. Connect port 1 to port 2, as shown inFigure 3-7. Figure 3-7 Port 1 and Port 2 ¼ Wavelength Load “Offset Load”. ¼ Wavelength Short “Offset Short”. X, P, and K 11644A...
User Information Changing the ¼ Wavelength Shim Calibration Definition Changing the ¼ Wavelength Shim Calibration Definition The calibration kit definition data provided with this kit has a nominal value for the ¼ wavelength shim offset delay. You may use the nominal value provided, or measure the exact thickness of the shim and use that value to calculate its exact offset delay.
Page 24
User Information Changing the ¼ Wavelength Shim Calibration Definition X, P, and K 11644A...
Use, Maintenance, and Care of the Devices Electrostatic Discharge Electrostatic Discharge Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets). Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged.
Use, Maintenance, and Care of the Devices Visual Inspection Figure 4-1 ESD Protection Setup Visual Inspection Visual inspection and, if necessary, cleaning should be done every time a connection is made. Inspect mating surfaces for dirt, dust, foreign particles, or scratches, which can degrade device performance.
Page 28
Use, Maintenance, and Care of the Devices Cleaning the Mating Plane Surfaces These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information. WARNING Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container.
Use, Maintenance, and Care of the Devices Connections Connections Good connections require a skilled operator. Slight errors in operator technique can have a significant effect on measurements and measurement uncertainties. The most common cause of measurement error is poor connections. The following procedures illustrate how to make good connections.
Use, Maintenance, and Care of the Devices Connections Aligning Two Precision Flanges 1. Place the slip pins in the top and bottom holes of one flange, as shown in Figure 4-3. Figure 4-3 Inserting Slip Pins 2. Using the pins as guides, carefully align the flanges and insert two screws in the diagonal corner holes, as shown in Figure 4-4.
Use, Maintenance, and Care of the Devices Connections Aligning a Precision and a Non-Precision Flange 1. Place an alignment pin (with head) in the corner hole of one flange. Place a second alignment pin in the diagonal corner hole of the second flange, as shown in Figure 4-5.
Use, Maintenance, and Care of the Devices Connections Tightening a Flange Connection NOTE The best connection has symmetrical pressure applied as you gradually tighten the screws. 1. In an “X” pattern (for equal compression), tighten all four screws using a hex ball driver. Do not over-tighten.
Use, Maintenance, and Care of the Devices Handling and Storage Handling and Storage • Install the protective end caps and store the calibration devices in the foam-lined storage case when not in use. • Never store devices loose in a box, or in a desk or bench drawer. This is the most common cause of device damage during storage.
Page 34
Use, Maintenance, and Care of the Devices Handling and Storage 4-10 X, P, and K 11644A...
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Agilent Technologies. X, P, and K 11644A...
A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. Refer to “Contacting Agilent” on page 6-4 for a list of offices.
Performance Verification Performance Test Performance Test Termination Return Loss Measurement For your convenience the following procedures have been included in this kit. Use this test, as needed, to check the performance of the terminations in this kit. Figure 5-1 Return Loss Test Setup 1.
Performance Verification Performance Test 4. Set the averaging factor to 512. 5. At the adapter test port, perform a 2–port TRL calibration. 6. Turn on the calibration. 7. Connect the termination you wish to test to port 1. 8. Measure the return loss (S ) of the load.
Performance Verification System Operation Checks System Operation Checks Use the following procedures and the standard section in this kit to verify the operation of your calibrated system. NOTE If you are using a different analyzer, refer to its documentation for specific measurement instructions.
Page 41
Performance Verification System Operation Checks 7. To set the display: a. Press b. Press RESPONSE [SCALE] [.1] [x1] c. Select [REF POSN] [10] [x1] d. Select [REF VALUE] [0] [x1] e. Press MEASUREMENT [RESTART] 8. After one complete measurement sweep, the displayed trace should look similar to the typical trace shown in Figure 5-3.
Performance Verification System Operation Checks The S and S of a Thru Measurement 1. Verify that the calibration is still on. 2. Verify that the averaging factor is set to 512. 3. Connect the standard section to ports 1 and 2 to form a thru connection. 4.
Performance Verification System Operation Checks The S and S of a Thru Measurement 1. Verify that the calibration is still on. 2. Verify that the averaging factor is set to 512. 3. Connect the standard section to ports 1 and 2 to form a thru connection. 4.
Performance Verification System Operation Checks In Case of Failure If a system check fails, recalibrate the system and repeat the entire procedure (all three measurements). Remember that poor connections are the most common cause of measurement errors. Also, both the hardware state and the instrument state must be correct.
Page 45
Performance Verification System Operation Checks Table 5-1 Using a TRL Calibration X11644A WR-90 (X-band) Test Measured Acceptable Pass Value Value (Yes/No) ≤ 0.15 dB , shorted ≤ 0.15 dB , shorted ≤ −45 dB , thru ≤ −45 dB , thru ≤...
Performance Verification System Operation Checks 872x System Operation Check 1. Short the standard section and perform an S and an S measurement to verify that the ripple on the reflection of the shorted waveguide section does not exceed the limits listed in the tables on page 5-15.
Performance Verification System Operation Checks 9. Record the maximum peak-to-peak value in the appropriate table on page 5-15. 10. Turn off the delta marker mode. 11. Repeat this procedure from step 6, and make an S measurement on the standard section.
Performance Verification System Operation Checks 7. Record the maximum peak-to-peak value in the appropriate table on page 5-15. 8. Repeat this procedure from step 3, and make an S measurement. Record the measured value in the appropriate table on page 5-15.
Page 49
Performance Verification System Operation Checks Table 5-4 Using a Full 2-Port Calibration X11644A WR-90 (X-band) Test Measured Acceptable Pass Value Value (Yes/No) ≤ 0.15 dB , shorted ≤ 0.15 dB , shorted ≤ −40 dB , thru ≤ −40 dB , thru ≤...
Page 50
Performance Verification System Operation Checks 5-16 X, P, and K 11644A...
Troubleshooting Troubleshooting Process Troubleshooting Process This manual contains limited information about network analyzer system operation. For complete information, refer to the instrument documentation. If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 6-1.
Troubleshooting Returning a Kit or Device to Agilent Returning a Kit or Device to Agilent If your kit or device requires service, contact Agilent Technologies for information on where to send it. Refer to “Contacting Agilent” on page 6-4. Include a service tag (located near the end of this manual) on which you provide the following information: •...
Troubleshooting Contacting Agilent Contacting Agilent Assistance with test and measurement needs and information on finding a local Agilent office are available on the Web at: http://www.agilent.com/find/assist If you do not have access to the Internet, please contact your Agilent field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number.
To order a listed part, note the description, part number, and the quantity desired. Telephone or send your order to Agilent Technologies. Refer to “Contacting Agilent” on page 6-4 for further information.
Page 58
Replaceable Parts Introduction Table 7-1 Replaceable Parts for the X11644A WR-90 Description Agilent Per Kit Part Number Calibration Devices Termination 00910-60003 Standard section 00896-60008 Short 11644-20018 ¼ Wavelength shim 11644-20021 7 mm coax-to-waveguide adapter X281C Option 006 Hardware Alignment pin 11644-20024 Slip pin 11644-20025...
Page 59
Replaceable Parts Introduction Figure 7-1 X-Band Component Identification Sheet X, P, and K 11644A...
Page 60
Replaceable Parts Introduction Table 7-2 Replaceable Parts for the P11644A WR-62 Agilent Description Per Kit Part Number Calibration Devices Termination 00910-60002 Standard section 00896-60007 Short 11644-20017 ¼ Wavelength shim 11644-20020 7 mm coax-to-waveguide adapter P281C Option 006 Hardware Alignment pin 11644-20023 Slip pin 11644-20025...
Page 61
Replaceable Parts Introduction Figure 7-2 P-Band Component Identification Sheet X, P, and K 11644A...
Page 62
Replaceable Parts Introduction Table 7-3 Replaceable Parts for the K11644A WR-42 Description Agilent Per Kit Part Number Calibration Devices Termination 00910-60001 Standard section 00896-60006 Short 11644-20016 ¼ Wavelength shim 11644-20019 3.5 mm coax-to-waveguide adapter (m) 00281-60001 3.5 mm coax-to-waveguide adapter (f) K281C Option 006 Hardware Alignment pin...
Page 63
Replaceable Parts Introduction Figure 7-3 K-Band Component Identification Sheet X, P, and K 11644A...
Page 64
Replaceable Parts Introduction Table 7-4 Items Not Included in the Calibration Kit Description Agilent Part Number ESD Protection Devices Grounding wrist strap 9300-1367 5 ft grounding cord for wrist strap 9300-0980 2 ft by 4 ft conductive table mat with 15 ft grounding 9300-0797 wire ESD heel strap...
Page 65
Standard Definitions X, P, and K 11644A...
Standard Definitions Version Changes Version Changes Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Agilent’s Calibration Kit Definitions Web page at http://na.tm.agilent.com/pna/caldefs/stddefs.html.
Standard Definitions Standard Class Assignments Standard Class Assignments Class assignment organizes calibration standards into a format compatible with the error models used in the measurement calibration. A class or group of classes corresponds to the systematic errors to be removed from the measured network analyzer response. Tables A-1 list the classes of the devices in the kit for the 8510 and PNA series respectively.
Page 68
Standard Definitions Standard Class Assignments Table A-1 Standard Class Assignments for the 8510 Network Analyzer Disk File Name: CK_WR-90, Calibration Kit Label: WR-90 A.1, or or CK_WR-62, or CK_WR-42 WR-62 A.1, or WR-42 A.1 Class Standard Class Label Short Offset Loads Short Offset...
Page 69
Standard Definitions Standard Class Assignments Table A-2 Standard Class Assignments for the PNA Series Network Analyzer SOLT Class TRL Class Assignments Assignments Calibration Kit Name: Calibration Kit Name: X11644A, P11644A, X11644A, P11644A, K11644A K11644A Class Class TRL thru TRL reflect 1, 2 7, 2 line/match...
Standard Definitions Standard Class Assignments Blank Forms The standard class assignments may be changed to meet your specific requirements. Tables A-3 are provided to record the modified standard class assignments. Table A-3 Blank Form for the 8510 Network Analyzer Disk File Name: _____________ Calibration Kit Label: _____________ Class Standard...
Page 71
Standard Definitions Standard Class Assignments Table A-4 Blank Form for the PNA Series Network Analyzer SOLT Class TRL Class Assignments Assignments Calibration Kit Name: Calibration Kit Name: ____________ ____________ Class Class TRL thru TRL reflect line/match Forward Isolation transmission Cal Ref Z0 = Line Z0 Isolation Transport Ref Plane = Thru Standard...
Standard Definitions Nominal Standard Definitions Nominal Standard Definitions Standard definitions provide the definitions needed to mathematically model the electrical characteristics (delay, attenuation, and impedance) of each calibration standard. The nominal values of these definitions are theoretically derived from the physical dimensions and material of each calibration standard, or from actual measured response.
Page 73
Standard Definitions Nominal Standard Definitions Table A-5 Standard Definitions for the 8510 with X-band (WR-90) = 1 Ω Calibration Kit Label: WR-90 A.1 System Z Disk File Name: CK_WR-90 Frequency Offset Standard in GHz 6.555 13.111 Short Short λ/4 Offset Short/ 32.633 6.555 13.111...
Page 74
Standard Definitions Nominal Standard Definitions Table A-6 Standard Definitions for the 8510 with P-band (WR-62) = 1 Ω Calibration Kit Label: WR-62 A.1 System Z Disk File Name: CK_WR-62 Frequency Offset Standard in GHz 9.485 18.97 Short Short λ/4 Offset Short/ 21.689 9.485 18.97...
Page 75
Standard Definitions Nominal Standard Definitions Table A-7 Standard Definitions for the 8510 with K-band (WR-42) = 1 Ω Calibration Kit Label: WR-42 A.1 System Z Disk File Name: CK_WR-42 Frequency Offset Standard in GHz 14.047 28.094 Short Short λ/4 Offset Short/ 15.015 14.047 28.094...
Page 76
Standard Definitions Nominal Standard Definitions Table A-8 Standard Definitions for the PNA Series with X-band (WR-90) = 1 Ω Calibration Kit Label: X-band System Z Frequency Offset Standard in GHz Offset Offset 6.555 13.111 Offset load load Load Fixed 6.555 13.111 Fixed Load Short...
Page 77
Standard Definitions Nominal Standard Definitions Table A-9 Standard Definitions for the PNA Series with P-Band (WR-62) = 1 Ω Calibration Kit Label: P-band System Z Frequency Offset Standard in GHz Offset Offset 9.485 18.970 Offset load load Load Fixed 9.485 18.970 Fixed Load Short...
Page 78
Standard Definitions Nominal Standard Definitions Table A-10 Standard Definitions for the PNA Series with K-Band (WR-42) = 1 Ω Calibration Kit Label: K-band System Z Frequency Offset Standard in GHz Offset Offset 14.047 28.094 Offset load load Load Fixed 14.047 28.094 Fixed Load Short 14.047 28.094...
Standard Definitions Nominal Standard Definitions Blank Forms The standard definitions may be changed to meet your specific requirements. Tables A-11 A-12 are provided to record the modified standard definitions. Table A-11 Blank Form for the 8510 Network Analyzer Calibration Kit Label: ____________________________ System Z = _________ ____________...
Page 80
Standard Definitions Nominal Standard Definitions Table A-12 Blank Form for the PNA Series Network Analyzer Calibration Kit Label: K-band __________________ System Z = ____________ Frequency Offset Standard in GHz a. Ensure system Z of network analyzer is set to this value. b.
Page 81
7-10 calibration Agilent Technologies performance cleaning contacting how Agilent verifies compatible analyzers Agilent Technologies application verifying connecting note case, storage handling alcohol, isopropyl certificate of calibration maintenance part number 7-10 certification, device specifications...
Page 82
Index return loss specification standard sections handling National Institute of Standards terminations heel strap, part number 7-10 and Technology (NIST) history, serial numbers humidity specifications network analyzers, compatible serial numbers nitrogen, for cleaning devices nominal standard definitions history impedance, system non-precision flanges, aligning recording incoming inspection...
Page 83
Index National Institute of Standards part number and Technology (NIST) wrist strap for ESD protection static electricity part number 7-10 storage storage case, part number strap heel and wrist, part number 7-10 supplies, cleaning part number 7-10 system impedance table mat for ESD protection part number 7-10...