To remove and replace the scope interface board and SVGA display board

When necessary, refer to other removal procedures.
Disconnect the power cable and remove the top cover.
1
Disconnect these cables from the SVGA display board:
2
• Bridge cable W22
• Backlight primary cable W21
• Flat panel display driver cable W11
Remove the two Torx T15 screws that secure the scope interface and SVGA display boards
3
to the chassis.
These screws are at the rear of the chassis.
Figure 5-25
Bridge
Cable
W22
Removing the Scope Interface and SVGA Display Boards
Grasp the scope interface board at the top corners and pull the board straight up until it is
4
free of the card cage.
Grasp the SVGA display board at the top corners and pull the board straight up until it is
5
free of the card cage.
To replace the scope interface and SVGA display boards, reverse the above procedure.
6
Be sure to observe correct polarity on all cables when replacing the boards.