Section 8
8.5

CALIBRATING THE THICKNESS GAUGE

If you purchased the 1011TS thickness transducer separately then calibrate the transducer before
using it to measure a pipe wall. The 1011TS Transducer Holder includes two built-in steel calibration
blocks. One is 0.100" thick and the other is 0.500" thick (see below). You will use these blocks in
conjunction with the Thickness Gauge Calibration menu to calibrate your transducer.
Transducer
Holder
Label
0.100
Identification
0.100
Calibration
Block
To calibrate the 1011TS transducer, remove it from its holder:
Siemens
CONTROLOTRON S/N
1011TS
0.100
THICKNESS GAUGE
Unscrew the Transducer Holder
Cap (counter-clockwise)
Remove the tensioning
spring
Withdraw the
transducer & cable
from the holder
These should be
connected according to
previous instructions
Transducer Holder
8-4
0.500
Identification
0.500
0.500
Calibration
Block
1010PFM-3G