Section 8
Holding the transducer against the calibration block, press
The thickness screen shows the 0.100 measurement (similar to figure below):
MENU
Press the
key to return to the calibration menu. Use the
Material], then press the
ber: [=0.500].
Now you have to place the 1011TS transducer flush on to the [0.500] calibration block on the trans-
ducer holder. Press
ENT
The thickness screen shows the 0.500 measurement (similar to figure below):
Siemens
MENU
Press the
key to return to the menu. This completes the calibration procedure. The 1011TS thick-
ness gauge transducer is now certified for operation with the flow computer.
NOTES
Repeat the calibration procedure for best accuracy.
The thickness screen shows the 0.500 measurement (similar to figure above). This cali-
brates the 1011TS satisfactorily for most pipe materials. If you have reason to suspect its
accuracy on your material, try the calibration procedure using a sample of your pipe material
of a known thickness.
If your pipe material is not listed in the "select material" menu, then refer to paragraph 8.6
"Manual Calibration of Thickness Gauge."
Siemens
to enable editing. Note that an equal (=) sign appears before the num-
Set Xdcr Zero
Select Material
Calibrate Material
to set the [Calibrate Material] value.
ENT
to calibrate the [Set Xdcr Zero] value.
to move the highlight to [Calibrate
0.100
Steel
=0.500
8-6
1010PFM-3G