Siemens SITRANS F Operating Instructions Manual page 77

Ultrasonic flowmeters portable flowmeter
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Section 1
NOTE: You are allowed to override the transducer type and size mounting mode and spacing
offset. See paragraph 1.9.
Now you have to mount the transducers in accordance with the data recommended on the screen.
Refer to the Section 3 for detailed instructions. The pipe must be completely filled with a liquid for the
initial installation of the meter. Make sure that you do not install your transducers backwards or reverse
the Upstream and Downstream cables.
Use the [Install Completed?] menu cell to inform the meter that you have finished mounting
the hardware. Press
The meter performs an Initial Makeup which conditions its operation to the selected transducer, pipe
and liquid and then sets the empty pipe and the zero flow levels automatically. Be patient! The Initial
Makeup performs a number of operations including a frequency sweep of the pipe wall resonance,
which could take several minutes.
Siemens
Siemens
. Move the cursor to [Install]. Press
Siemens
.
ENT
1-33
1010PFM-3G