Output Datalogger; Circular Memory - Siemens SITRANS F Operating Instructions Manual

Ultrasonic flowmeters portable flowmeter
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Section 4
To view Datalogger contents press
To return to [Datalogger Control] press

4.4.2 OUTPUT DATALOGGER

This menu cell allows you to send the Datalogger contents to an external device (usually a computer or
printer) via the meter's RS-232 Serial I/O port. This command is effective only after a successful
install. In addition, you have to enable datalogging and select data items in the Datalogger Setup menu.
Note: This command sends the data collected by both channels of a dual-channel system. Therefore,
you should include [Site ID] (i.e., current site setup name), [Date] and [Time] in your report so that you
can identify the source of each report.
The meter interfaces with most serial printers or personal computers. Controlotron offers the Model
996P portable serial printer. You must use the proper cabling between the flow computer and the
external device. In addition, you must configure the RS-232 Setup correctly. You should turn off the
Datalogger function before you transmit an extensive printout. This will avoid contaminating the printout
with new Datalogger data. Datalogger reports are sequential ASCII text files.
To send Datalogger contents to the RS-232 Serial Port:
Check the flow computer-to-external device connections and your RS-232 Setup para-
meters (see RS-232 Setup).
To access the [Output Datalogger] option list press
Scroll the cursor to [Yes] press
To transmit Datalogger contents to external device via the serial port press
To stop printout press

4.4.3 CIRCULAR MEMORY

In its default mode, the Datalogger collects data until its memory becomes full. At that time the flowme-
ter suspends data logging and cannot resume until the datalogger memory is cleared (see Clear
Datalogger command). Multi-Channel systems include Circular Memory and Single Channel systems
do not. Circular Memory allows the Datalogger to "'write over" its oldest records when memory reaches
full capacity. If you enable Circular Memory, you are assured of always collecting the most recent data.
But, also remember that you will lose the oldest Datalogger reports. Note further that invoking Circular
Memory deletes the current contents of the Datalogger.
NOTE: To avoid potential data loss, the 1010 flowmeter prevents you from altering the Circu-
lar Memory setting when a channel is active.
To setup and enable Circular Memory:
The Datalogger Mode menu must have the [Memory] menu cell selected.
Datalogger items must be selected (e.g., Site ID, Date, Time, etc.).
All active channels in the Channel Setup menu must be disabled. To disable active channels,
select the [Channel Enable] menu cell and then [No].
ENT
MENU
4-7
1010PFM-3G
ENT