Section 3
We recommend Reflect mount whenever possible. This is the simplest way to mount the transducers.
Also, Reflect mount resists abnormal flow profile conditions such as cross-flow within the flow stream.
Reflect mode supports the AutoZero function, which zeroes the meter automatically without user-
participation. In addition, Reflect mount may be the only possibility if conditions do not allow access to
the opposite side of the pipe.
Direct mount provides a shorter sonic beam path. This usually improves performance with sonically
attenuative liquids or pipe materials. We recommend using Direct mount for plastic pipes. Compared
to Direct mounting, Reflect mount requires almost double the amount of a straight pipe run. Therefore,
Direct mount may be the only option if the availability of mounting space is limited.
Upstream
Upstream
Transducer
Transducer
Sonic Path
Sonic Path
(Pipe shown in Top View for ease in visualizing sonic path)

3.1.4 PREPARING THE PIPE

Pick a mounting location with the longest straight run. You must have easy access to at least one
side of your pipe. The mounting location must remain full, even at zero flow.
Decide on your mounting mode (Direct or Reflect). Always use Reflect Mode whenever possible.
You may only need to use Direct Mode if your pipe is plastic.
After receiving the spacing dimensions from the Installation Menu, prepare the pipe surface. De-
grease the surface, if necessary, and remove any grit, corrosion, rust, loose paint, etc. Use abra-
sive material provided to provide a clean contact surface for the transducers.
Refer to the next sections for illustrated instructions on how to locate each area to be cleaned and
how to use each mounting option.
Please note that the instructions show vertical mounting for clarity purposes only. Do not
install transducers on the top of a pipe.
PREPARING THE PIPE SURFACE
Downstream
Downstream
Transducer
Transducer
DIRECT MOUNT
Pipe
1 2 3 4 5 6 7 8 9 0
1 2 3 4 5 6 7 8 9 0
1 2 3 4 5 6 7 8 9 0
1 2 3 4 5 6 7 8 9 0
1 2 3 4 5 6 7 8 9 0
1 2 3 4 5 6 7 8 9 0
3-3
Pipe
Pipe
Cleaned areas. Clean an area 1/2" on either side
of transducer and an additional 1/2" along the
length of the transducer.
1010PFM-3G