Section 2
Press <ENT>. The [Install Completed?] menu cell will indicate [Yes] after the 1010 is successfully
installed.
To complete the Install process, disregard the following paragraphs explaining the Force Trans-
mit and Force Frequency diagnostic software routines.
NOTE: If the transducers have been installed successfully but the Estimated Vs (sonic veloc-
ity) has been changed, the Pick/Install Xdcr menu cell [Install Completed?] option list
will also display the [Transfer Install] function selection. The Transfer Install function
allows the transducers to be optimally positioned for a different fluid, without the need
for a new Initial Makeup procedure (see the Water Calibration addendum in this manual
for details).

Force Transmit Procedure

This diagnostic software routine allows the user to "force" a transmitting condition that can be use to
search for an amplitude level (ALC) when Detection Fault or Low Signal alarms are present. The
routine forces the flowmeter to generate constant transmit bursts while reporting current receive signal
strength for the user. To initiate the Force Transmit function, refer to the example shown below.
1. After [Install] is selected press <ENT>. While the meter is going through the drives (see menu
screen below), press the <ALT> and <MENU> keys simultaneously.
Siemens
2 Channel
Key [Install] after mounting transducers
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value (in)
Install Completed?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
Siemens
2 Channel
Key [Install] after mounting transducers
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value (in)
Install Completed?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
[1]
SITE1
1011 Universal
B3
Direct
Minimum
4
Track 1012TP
0.217
No
Channel Not Setup
Channel Not Setup
2-22
[1]
SITE1
1011 Universal
B3
Direct
Minimum
(631) 248 77774
Track 1012TP
0.217
Yes
Channel Not Setup
Channel Not Setup
Siemens
2 Channel
Drive 11
Transducer Model
Transducer Size
Xdcr Mount Mode
Spacing Offset
Number Index
Spacing Method
Ltn Value (in)
Install Completed?
Empty Pipe Set
Zero Flow Adjust
Pick/Install Xdcr
1010PFM-3G
[1]
SITE1
1011 Universal
B3
Direct
Minimum
4
Track 1012TP
0.217
Install
Channel Not Setup
Channel Not Setup