Section 2
Xdcr
Clamping
Bracket
To mount Transducers on a 996PSP Pipe Simulator:
Using coupling compound, mount the transducers on the pipe simulator as shown on the previous
page. Slide each transducer until it presses against a pin-stop. Use the clamps to hold the trans-
ducer in place.
Connect cables between each transducer and the computer connectors for the channel under
test. The Up and Down orientation is not important.
Access the Installation Menu. Select [Meter Type] [Single, Dual, or Quad Channel] depending on
meter type. Select the meter channel (1,2,3, or 4) depending on which measurement channel you
intend to test. Select [Clamp-On] and then [Channel Setup].
Access [Channel/Path Setup] menu. Move highlight to [Create/Name Site]. Create a new Site Setup
(e.g., TEST1). You can now enter data without altering an existing Site Setup.
Select the [Pipe Data] Menu. Referencing either the English or metric pipe simulator chart below,
enter the pipe parameters corresponding to the transducer size under test.
Part
Number
996PSP-0
996PSP-1
996PSP-2
996PSP-3
996PSP-4
* The computer generates the Number Index.
** See note below.
NOTE: When using a 996PS-4 Simulator, you must access the [Application Data] menu, select
[Liquid Type] - [Other], then program the [Estimated Vs] for a value of 2100 m/s.
Clamping Screw
Pipe Simulator
996PSP PIPE SIMULATOR CHART (English)
Xdcr
Pipe
Pipe
Size
OD (in)
Mat'l
0
2.420
PVC
1
2.481
PVC
2
2.743
PVC
3
3.758
PVC
4
6.500
PVC
Transducer
Pinstops
Wall
Mount
Letter
Thk (in)
Mode
0.076
Direct
0.077
Direct
0.136
Direct
0.344
Direct
1.000
Direct**
2-77
1010PFM-3G
991 Transducers
Under test
Number
Index
Index*
A
6
A
9
A
6
A
1
A
0