Section 6
6.6
SPECIFICATIONS – TEFLON® PFA FLOW TUBE
*Typical values, subject to change without notice
Liquid
Pressure Drop
Sensitivity
Linearity
Repeatability
Zero Stability
Accuracy,
Intrinsic
Accuracy, Flow
Calibrated
Connection
6.7
SPECIFICATIONS – 316 STAINLESS STEEL FLOW TUBE
*Subject to change without notice
Liquid
Pressure Drop
Sensitivity
Linearity
Repeatability
Zero Stability
Accuracy,
Intrinsic
Accuracy, Flow
Calibrated
Connection
Size 1
Must be compatible with PFA under
operating conditions and free of sus-
pended gasses or solids.
0.23 psi @ 1.0 GPM (3.8 LPM)
flowing water at 68°F
0.0004 GPM (0.0015 LPM)
1% of flow for rates above 0.07 GPM
(0.26 LPM)
±0.2% of flow for rates above 0.6
GPM (2.3 LPM)
±0.001 GPM (0.0038 LPM)
±1% of flow over 0.15 GPM (0.57
LPM) to 5 GPM (20 LPM)
±0.5% of flow over 0.15 GPM (0.57
LPM) to 5 GPM (20 LPM)
1
"
/
Flaretek Ultrapure PFA Tube Fitting
2
Size 1
Must be compatible with 316SS
under operating conditions and free
of suspended gasses or solids.
0.21 psi @ 1.6 GPM (6.0 LPM) flow-
ing water at 68°F
0.0005 GPM (0.002 LPM)
1% of flow for rates above 0.1 GPM
(0.38 LPM)
±0.2% of flow for rates above 0.5
GPM (1.9 LPM)
±0.001 GPM (0.0038 LPM)
±1% of flow over 0.13 GPM (0.57
LPM) to 10 GPM (40 LPM)
±0.5% of flow over 0.13 GPM (0.5
LPM) to 10 GPM (38 LPM)
5
/
" Swagelok Tube Fitting
8
Must be compatible with PFA under
operating conditions and free of
suspended gasses or solids.
0.23 psi @ 3.0 GPM (11.4 LPM)
flowing water at 68°F
0.001 GPM (0.004 LPM)
1% of flow for rates above 0.27 GPM
(1.00 LPM)
±0.2% of flow for rates above 1.5 GPM
(5.7 LPM)
±0.003 GPM (0.012 LPM)
±1% of flow over 0.5 GPM (1.9 LPM) to
16 GPM (60 LPM)
±0.5% of flow over 0.5 GPM (1.9 LPM)
to 16 GPM (60 LPM)
3
/
" Flaretek Ultrapure PFA Tube Fitting
4
Must be compatible with 316SS under
operating conditions and free of sus-
pended gasses or solids.
0.20 psi @ 4.5 GPM (17.0 LPM)
flowing water at 68°F
0.0015 GPM (0.0057 LPM)
1% of flow for rates above 0.3 GPM
(1.1 LPM)
±0.2% of flow for rates above 1.5
GPM (5.7 LPM)
±0.004 GPM (0.015 LPM)
±1% of flow over 0.4 GPM (1.5
LPM) to 30 GPM (114 LPM)
±0.5% of flow over 0.4 GPM (1.5
LPM) to 30(114 LPM)
1" Swagelok Tube Fitting
6-10
1010PFM-3G
Size 2
Size 2