Large Side Border from General Electric
SENOGRAPHE ACQUISITION WORKSTATION
Problem
Images that are acquired from the General Electric SENOGRAPHE 2000D FFDM ACQUISITION
WORKSTATION are printed smaller than the images printed from the General Electric REVIEW
WORKSTATION. This results in a larger border on one side of the image.
Cause
General Electric adds patient information to the bottom of the images that are printed from the
ACQUISITION WORKSTATION. This patient information is not included on images printed from the
REVIEW WORKSTATION.
The patient information results in a shorter image height and to keep the correct aspect ratio, the image
width is also cut down. The smaller width increases the border on one side of the image.
Solution
Inform the customer that this is a function of the ACQUISITION WORKSTATION and that images printed
from the REVIEW WORKSTATION will be true size, but images printed from the ACQUISITION
WORKSTATION will not.