NOTE: Do not touch the MIRROR with the NOZZLE and do not tilt, invert or shake the CANISTER
while spraying air into the opening. A tilted CANISTER will cause fluid to be sprayed which will
permanently damage the MIRROR.
5. Remove the NOZZLE and install the OPTICS MODULE back into the IMAGER.
6. Make another print of a film calibration sheet.
7. Compare the print with the film calibration sheet from Step 1 to confirm that the light streaks have
been eliminated.