Technical Bulletin
17-09/01
Light Density Streaks on Film
Problem
Light streaks appear down the film, usually around the middle, and most are visible in the mid-tone grays
of the images.
Cause
Dust has collected on the final MIRROR that directs the laser beam to the film.
Solution
1. Print an image to be used for comparison. A film calibration sheet might show the artifact.
2. Remove the OPTICS MODULE from the imager and set it on a TABLE in the same position as when
it is in the imager.
3. Insert the NOZZLE from the CANISTER in the OPTICS MODULE CLEANING KIT
74-0401-82380-3 into the laser-beam aperture slot toward the top of the OPTICS MODULE.
4. Holding the CANISTER upright, spray short bursts while moving the NOZZLE back and forth along
the aperture slot.
Note
Do not touch the MIRROR with the NOZZLE and do not tilt, invert or shake the CANISTER while
spraying air into the opening.
5. Remove the NOZZLE and replace the OPTICS MODULE into the imager.
6. Run a comparison image to check that the light streak has been removed.
September 2001