"Fail to Get DICOM Printer Status" Modality Error
Message
Problem: A new installation with a GE FFDM system passes successful Pings and Send Echo
commands, but the error message "Fail to Get DICOM Printer Status" displays on the modality's
monitor when a Print command is sent to the 8600/9410.
Cause: The wrong modality AE Title was entered for the Registered User on the 9410. It was
originally documented that "Print_SCU" was the default AE Title for GE Senographe 2000D. However,
this is not always the case.
Solution: Obtain the proper AE Title from the GE representative and enter it for the Registered User
in the SCP Printer file.
2-1/00
QC Step Wedge Density Measurements Not as
Expected
Problem: The density measurements of specific steps in the QC step wedge are not what the OEM
expects for the (Contrast) Modality Group selected in the 8600 or 9410.
Cause: The selected Modality Group is not the correct group of contrast tables (those recommended
by the OEM). The groups have been renumbered since you first received information on group
numbering in the original 8600 Training Bulletin. The Modality Group numbers now range from 1-11
instead of 0-10 as in the past.
Solution: Change the Modality Group number to reflect the desired group. In most cases, this is as
simple as increasing the number by one. For example, if group 5 was used when the groups were
labeled 0-10, group 6 will need to be selected now that the groups are labeled 1-11.