Kodak DryView 8300 Service Bulletins page 35

Laser imaging system
Hide thumbs Also See for Kodak DryView 8300:
Table of Contents

Advertisement

Service & Support
Health Imaging
Kodak DryView 8300 LASER IMAGERS
Technical Bulletin
15-10/01
Cost Savings for Replacing PRESSURE ROLLERS
If the UPPER PRESSURE ROLLERS in the PROCESSOR stop turning, the ROLLERS can deplate or
flatten. If this occurs, the DRUM will be covered with aluminum deposits and change to a black color. If
you have a problem with the UPPER PRESSURE ROLLERS, installing only the ROLLER ASSEMBLY in
place of installing a new PROCESSOR ASSEMBLY will result in a cost savings for Kodak.
To correct this problem, install a new ROLLER ASSEMBLY(S) and clean the DRUM. This will result in
the following cost savings:
Installing 9 ROLLER ASSEMBLY (cover) and 3 ROLLER ASSEMBLY (base):
savings = $2506.00.
Installing 9 ROLLER ASSEMBLY (cover), 3 ROLLER ASSEMBLY (cover), and the DRUM:
savings = $1651.00.
October 2001