Kodak DryView 8300 Service Bulletins page 116

Laser imaging system
Hide thumbs Also See for Kodak DryView 8300:
Table of Contents

Advertisement

SERVICE BULLETIN
Health Imaging Products
Eastman Kodak Company, Health Imaging, Rochester, NY 14650
SERVICE BULLETIN NO. 37
October 2002
1437 and 1439 Kodak DryView 8600/8610 LASER IMAGERS
EC 67 FIFO Not Empty Failure When Connected to the Kodak PACS LINK MEDICAL
IMAGE MANAGER 200 (MIM 200)
Purpose
The purpose of this service bulletin is to communicate the cause of the EC 67, when the PRINT
SERVER for the IMAGER is a MIM 200.
Problem
EC 67 is generated when the FIFO does not empty in the DIGITAL INPUT MODULE of the
IMAGER.
Cause
The COPCIL BOARD in the MIM 200 is causing the IMAGER to incorrectly generate the error
message.
Solution
Replace the COPCIL in the MIM 200.
Eastman Kodak Company, 2002
Pub No. 8E6990

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Kodak dryview 8610Kodak dryview 8600

Table of Contents