Kodak DryView 8300 Service Bulletins page 12

Laser imaging system
Hide thumbs Also See for Kodak DryView 8300:
Table of Contents

Advertisement

SERVICE BULLETIN
Health Imaging Products
Eastman Kodak Company, Health Imaging, Rochester, NY 14650
SERVICE BULLETIN NO. 23
1425, 1426 and 1427 Kodak DryView 8300 LASER IMAGER
Instructions for Grounding the PROCESSOR ASSEMBLY to Eliminate EC88s
Purpose
The purpose of this service bulletin is to provide instructions for grounding the PROCESSOR ASSEMBLY
using the PROCESSOR GROUNDING HARDWARE KIT, 8E2003.
Procedure
1. Remove the SPRING CLIPS from the FRAME and CLAMSHELL of the PROCESSOR
ASSEMBLY (see Fig. 1).
2. Attach one end of the WIRE (supplied in the KIT) to the STRIPPER ASSEMBLY using the right
mounting SCREW of the STRIPPER ASSEMBLY (see Fig. 2).
3. Attach one end of 1 small STRAP (supplied in the KIT) along with the loose end of the WIRE
referred to in Step 2, to the FRAME where the CLIP was removed in Step 1 (see Fig. 2).
4. Remove the SCREW holding the ROLLER ASSEMBLY MOUNTING ROD, and using the supplied
SPACER and SCREW, attach the loose end of the small STRAP to the ROLLER ASSEMBLY
MOUNTING ROD (see Fig. 2).
5. Repeat Steps 3 and 4 to attach the remaining small STRAP to either side of the CLAMSHELL
portion of the PROCESSOR ASSEMBLY.
ãEastman Kodak Company, 2002
AUGUST 2002
Pub No. 8E6688