Maintenance
4
4.1
4.2
4.3
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 10
Agfa Company Confidential

Maintenance points - Power off

The maintenance with powered-off machine comprises following tasks:
Task

Check of general condition

Check of condition of belts and gears

Visual check of the cables

Cleaning of the suction cups

Cleaning of Electronics
Cleaning of Power Supply Air Inlet
Check and Cleaning of Drum
Cleaning of Thermal print head
Electrical check
Check of general condition
(1) Perform a visual check of
Power- and network cables.
Covers
The Output tray
The upper and lower Input tray.
(2) Disconnect the power and network cable
(3) Remove the left and right cover to prepare the printer for the next steps
(4) Remove the films in the film trays including white plastic sheet
Check of condition of belts and gears
(1) Check all belts and gears for visible wear
Visual check of the cables
(1) Check cables inside the printer: The insulation may not show any sign of
wear or damage.
Printed copies of this document are not controlled.
DRYSTAR 5302
(Type 5366/100)
DD+DIS149.05E
Details: see...
4.1
4.2
4.3
4.4
4.5
4.6
4.7
4.8
4.9
Edition 1, Revision 0
2005-09-09

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dd+dis150.05e

Table of Contents