AGFA DRYSTAR 5302 Technical Documentation Manual page 260

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150
.05E
REQUIRED TOOLS:
Service PC with network interface and browser Internet Explorer
Calibrated densitometer MacBeth TR-924, X-Rite 331, 341, X-Rite 301, X-Rite
310 or Gretag D200-2
Lint free cloth
Isopropyl - alcohol
Waterproof pen
Density reproduction check Excel sheets, available via MEDNET, GSO Library
Afssaps / EUREF Calibration Check Excel Sheet, available via MEDNET, GSO
Library
Connectivity Release Document for the corresponding modality to DRYSTAR
5302, available via MedNet HealthCare Library
Application
REQUIRED TIME:
60 to 120 min.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
General Info
Connectivity &
Chapter 3.6 / 18
Agfa Company Confidential