AGFA DRYSTAR 5302 Technical Documentation Manual page 221

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
HC314004 : gearboxmodule.m8.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC314005 : gearboxmodule.s3
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC314006 : gearboxmodule.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Defect
E314004
SeeManual
Motor M8 not connected
Connect motor M8
Roller drive module defect
Replace
MOD14
Defect
E314005
SeeManual
Sensor S3 not connected
Connect sensor S3
Roller drive module defect
Replace
MOD14
Defect
E314006
SeeManual
Connector XS12 not connected
Connect Connector XS12
gearboxmodule drive module defect
Replace
MOD14