AGFA DRYSTAR 5302 Technical Documentation Manual page 247

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
Instructions
Via Printer Browser User Interface
(1)
In the printer browser user interface select 'Service Engineer Tools'.
(2)
Select 'Enter Service Command Session'.
(3)
Select '6 Calibration - 4 Horizontal Alignment'.
(4)
Select 'upper tray' or 'lower tray': A test film as shown in Figure 1 will be
printed.
(5)
Measure the left and right blank border of the test film.
(6)
Enter the values in mm (typically approx. 10 mm) in the browser screen.
(7)
Select 'Accept Changes'.
Via Local Keypad
(1)
In the key-operator menu select menu '7 Service Actions'.
(2)
Select '6 Calibration - 4 Horizontal Alignment'.
(3)
Select 'upper tray' or 'lower tray': A test film as shown in Figure 1 will be
printed.
(4)
Measure the left and right blank border of the test film.
(5)
Enter the values in mm (typically approx. 10 mm) in the browser screen.
(6)
Select 'confirm'.
via local display
Measure left
white border
0.0 mm
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Left white border
Measure the left clear border width:
Value:
mm
Measure right
white border
0.0 mm
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Adjustments and Calibrations
Measure the right clear border width:
Right white border
Value:
mm
via Browser User Interface:
Welcome to this
DryStar 5302
Service - Calibration
Horizontal alignment
0.0
Left white border
0.0
Right white border
Accept changes
Discard changes
Figure 1
(mm)
(mm)
536736ac.cdr
Chapter 3.6 / 5
Agfa Company Confidential