Compliance Film Calibration As Part Of Afssaps / Euref Calibration Procedure - AGFA DRYSTAR 5302 Technical Documentation Manual

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
2.5
Compliance Film Calibration as part of Afssaps / EUREF Calibration
Procedure
Make film calibration this accurate, that it passes the acceptance criteria as defined
Purpose
by Afssaps / EUREF:
The default film calibration procedure is not necessarily accurate enough.
The compliance film calibration may be required as last step of the Afssaps / EUREF*
When to be
performed
calibration procedure.
The Afssaps / EUREF compliance calibration procedure has to be performed in
following cases:
Before the Afssaps / EUREF acceptance test
Each 6 months, if the printer is under control of Afssaps / EUREF, and it did not
pass the Afssaps / EUREF calibration check (see section 2.5.6).
*Afssaps
=
EUREF
=
1) The compliance film calibration is only
Prerequiste
available for DRYSTAR 5302
software ≥ 2.0.0_C1.
2) A "TG18-PQC modified 4" image has
to be available on the modality.
The Test Pattern can be downloaded
from http://www.euref.org/.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
Agence française de sécurité sanitaire des produits de santé
European Reference Organisation for Quality Assured Breast Screening and
Diagnostic Services
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Adjustments and Calibrations
Figure 10: TG18-PQC modified 4
Agfa Company Confidential
Chapter 3.6 / 17