Power Supply - AGFA DRYSTAR 5302 Technical Documentation Manual

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
3.3
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
2005-09-09
Electrical and mechanical codings, fuse tables and LED's

Power Supply

536634ab.cdr
LED
Status
Overtemp
On
Off
+ 3.3 V
On
Off
+ 5 V
On
Off
+12 V
On
Off
-12 V
On
Off
+24 V
On
Off
13..27 V
On
Off
*When the temperature inside the power supply exceeds the permissible limit,
printing can continue for up to 120 seconds. After this delay the power supply
switches off the power and checks the temperature every 20 seconds. As
soon as the temperature is below the permissible limit, it switches on again.
Printed copies of this document are not controlled.
Drystar 5302
(Type 5366/100)
Figure 3
Color
Function
Red
Temperature inside power supply
exceeded permissible limits. Operation
not possible*.
Temperature inside power supply
within permissible limit.
Green
+3.3 V present
+3.3 V not present
Green
+5 V present
+5 V not present
Green
+12 V present
+12 V not present
Green
-12 V present
-12 V not present
Green
+24 V present
+24 V not present
Green
voltage between 13 V and 27 V
voltage < 13 V
Repair and Service
Chapter 3.4 / 5
Agfa Company Confidential