DD+DIS238.06E
16

General Safety Directions for Service Activities

This system uses high voltage. Please consider the respective safety regulations.
Electrical repairs and connections must only be performed by a qualified electrician.
Mechanical repairs and connections must only be performed by a qualified technician.
The safety directions for operation (see section 13) are also valid for all service
activities.
During all service activities observe prescribed local and country-specific
requirements (e.g. occupational safety and accident prevention regulations).
All existing screw connections must be tightened sufficiently firmly, but they may not
be overstressed when tightening. There must always be compliance with stated
torque values!
Damaged or missing screws may be replaced only with the same screw types that
have the specified hardness rating. Unless a different value is listed in the
instructions, all Allen screws used must be hardness rated 8.8.
All screws must be secured in accordance with the corresponding data.
If "Loctite" has to be used to secure screws, this is stated in the text.
Any Agfa service PC or tool which is to be connected via RS232, RJ45, USB or other
interface to an Agfa device must not be connected to the mains but must be operated
on its internal battery or indirect supply (low voltage).
• When handling printed circuit boards (abbr.: PCBs) the following points must be
observed:
o
o
o
o
Static discharge at electrical components:
CAUTION:
Static discharge! Electrical components may be destroyed:
For the repair on electrical components, wear a grounding strap
(Order number: CM+9 9999 0830 0) around the wrist and connect the other end of this
strap on a grounded conducting metal piece.
Edition 1, Revision 3
07-2009
Always switch off the equipment and unplug the power cord, before you
disconnect or connect cables on printed circuit boards.
When working on PCBs, always wear an anti-static wrist strap. Never touch any
parts or components on PCBs with your bare fingers.
PCBs have to be kept or transported in their protection bags. Never carry a
PCB without protection bag and walk on carpet or plastic floor covering
(electrostatic charge).
Once the PCB is taken out of its protection bag, it has to be protected from
electrostatic charge by a grounded mat.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Generic Safety Directions for HealthCare Imaging Products
DOCUMENT CONTROL NOTE:
Generic Safety Directions
Page 19 of
Agfa Company Confidential
28