Welcome to this
DryStar 5302 printer
Service - Calibration
1.
Film Calibration
2.
Internal Densitometer Calibration
3.
Thermal Print Head Calibration
4.
Horizontal Alignment
5.
Smart profile calibration
Go back
Welcome to this
DryStar 5302 printer
Service - Calibration
Printhead Calibration
Upper Tray
Lower Tray
Field 21
DOCUMENT CONTROL NOTE:
Adjustments and Calibrations
Thermal Printhead Calibration
Start Calibration
Enter below the measurements of the printer calibration sheet
536736ag.cdr
Accept changes
Figure 4
DRYSTAR 5302
Type 5366/100
Repair and Service
Welcome to this
DryStar 5302 printer
Service - Calibration
Select Densitometer
Macbeth TR924
Gretag
X-Rite 301
X-Rite 310
Go back
Reset Values
Welcome to this
DryStar 5302 printer
Succeeded
Step
Step
2.0
2.1
00
01
2.0
2.1
02
03
2.1
2.1
04
05
2.1
2.1
06
07
2.2
2.1
08
09
2.1
2.2
10
11
2.1
2.1
12
13
2.1
2.0
14
15
2.1
2.1
16
17
2.1
2.1
18
19
2.1
2.2
20
21
Discard changes
Agfa Company Confidential
Reset values
Chapter 3.6 / 12