Software Release 1.9.1_C1; Solved Software Bugs - AGFA DRYSTAR 5302 Technical Documentation Manual

Hide thumbs Also See for DRYSTAR 5302:
Table of Contents

Advertisement

DD+DIS150.05E
7

Software Release 1.9.1_C1

Type of new Software*
Release Date
Printer Serial Numbers
Availability
Prerequisites
*NOTE:
 A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
 On MedNet, GSO Library always the latest available software release can be
downloaded.
7.1

Solved Software Bugs

Software 1.9.1_C2 solves following software bugs:
Symptom 1
Sporadic input film jams for films from the upper tray, mainly 14" films:
Error message at the local display:
Film jam! Open upper input tray and clear film path
Error message in the infocounter: HC314003
Cause
The force for driving the film towards the thermal head is sporadically too low.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2011
Software Releases and Patches
Software Upgrade
04/2008
≥ SN 5723 plus 5660, 5665, 5667, 5669,
5672, 5676, 5677, 5679, 5683, 5684, 5717,
5719, 5721
MedNet, GSO Library
None
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Type 5366/100
Repair and Service
Chapter 3.8 / 13
Agfa Company Confidential